Sustained safety and performance of the second-generation drug-eluting absorbable metal scaffold in patients with de novo coronary lesions: 12-month clinical results and angiographic findings of the BIOSOLVE-II first-in-man trial

نویسندگان

  • Michael Haude
  • Hüseyin Ince
  • Alexandre Abizaid
  • Ralph Toelg
  • Pedro Alves Lemos
  • Clemens von Birgelen
  • Evald Høj Christiansen
  • William Wijns
  • Franz-Josef Neumann
  • Christoph Kaiser
  • Eric Eeckhout
  • Soo Teik Lim
  • Javier Escaned
  • Yoshinobu Onuma
  • Hector M. Garcia-Garcia
  • Ron Waksman
چکیده

AIMS Metal absorbable scaffolds constitute a conceptually attractive alternative to polymeric scaffolds. Promising 6-month outcomes of a second-generation drug-eluting absorbable metal scaffold (DREAMS 2G), consisting of an absorbable magnesium scaffold backbone, have been reported. We assessed the 12-month safety and performance of this novel device. METHODS AND RESULTS The prospective, international, multi-centre, first-in-man BIOSOLVE-II trial enrolled 123 patients with up to two de novo lesions with a reference diameter between 2.2 and 3.7 mm. All patients were scheduled for angiographic follow-up at 6 months, and-if subjects consented-at 12 months. Dual antiplatelet therapy was recommended for 6 months. Quantitative coronary angiography (QCA) parameters remained stable from 6 to 12 months [paired data of 42 patients: in-segment late lumen loss 0.20 ± 0.21 mm vs. 0.25 ± 0.22 mm, P = 0.117, Δ 0.05 ± 0.21 mm (95% CI: -0.01;0.12); in-scaffold late lumen loss 0.37 ± 0.25 mm vs. 0.39 ± 0.27 mm, P = 0.446, Δ 0.03 ± 0.22 (95% CI: -0.04;0.10), respectively]. Intravascular ultrasound and optical coherence tomography findings corroborated the QCA results. Target lesion failure occurred in four patients (3.4%), consisting of one death of unknown cause, one target-vessel myocardial infarction, and two clinically driven target lesion revascularization. No additional event occurred beyond the 6-month follow-up. During the entire follow-up of 12 months, none of the patients experienced a definite or probable scaffold thrombosis. CONCLUSION The novel drug-eluting metal absorbable scaffold DREAMS 2G showed a continuous favourable safety profile up to 12 months and stable angiographic parameters between 6 and 12 months. CLINICALTRIALSGOV IDENTIFIER NCT01960504.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Ultra-low-dose hybrid single photon emission computed tomography and coronary computed tomography angiography: a comprehensive and non-invasive diagnostic workup of suspected coronary artery disease.

tion 2015;10:1160–1168. 13. Verheye S, Ormiston JA, Stewart J, Webster M, Sanidas E, Costa R, Costa JR Jr, Chamie D, Abizaid AS, Pinto I, Morrison L, Toyloy S, Bhat V, Yan J, Abizaid A. A next-generation bioresorbable coronary scaffold system: from bench to first clinical evaluation: 6and 12-month clinical and multimodality imaging results. JACC Cardiovasc Interv 2014;7:89–99. 14. Maeng M, Jens...

متن کامل

Serial observation of drug-eluting absorbable metal scaffold: multi-imaging modality assessment.

BACKGROUND The drug-eluting absorbable metal scaffold has demonstrated feasibility, safety, and promising clinical and angiographic outcomes at 12 months in human coronary arteries. This study aimed to evaluate the degradation rate and long-term vascular responses to drug-eluting absorbable metal scaffold. METHODS AND RESULTS BIOSOLVE-I was a multicenter, single-arm, first-in-man trial assess...

متن کامل

A randomized comparison of the Endeavor zotarolimus-eluting stent versus the TAXUS paclitaxel-eluting stent in de novo native coronary lesions 12-month outcomes from the ENDEAVOR IV trial.

OBJECTIVES The ENDEAVOR IV (Randomized Comparison of Zotarolimus-Eluting and Paclitaxel-Eluting Stents in Patients with Coronary Artery Disease) trial evaluated the safety and efficacy of the zotarolimus-eluting stent (ZES) compared with the paclitaxel-eluting stent (PES). BACKGROUND First-generation drug-eluting stents have reduced angiographic and clinical restenosis, but long-term safety r...

متن کامل

Comparison of a drug-eluting balloon first and then bare metal stent with a drug-eluting stent for treatment of de novo lesions: study protocol of a randomized controlled trial

BACKGROUND The use of a drug-eluting balloon for the treatment of de novo coronary artery lesions remains to be evaluated. A previous trial in patients with stable and unstable angina comparing a bare metal stent mounted on a drug-eluting balloon with a sirolimus-eluting stent failed to meet the prespecified non-inferiority criteria versus the sirolimus-eluting stent. The stent struts of a bare...

متن کامل

Serial Multimodality Imaging and 2-Year Clinical Outcomes of the Novel DESolve Novolimus-Eluting Bioresorbable Coronary Scaffold System for the Treatment of Single De Novo Coronary Lesions.

OBJECTIVES This study sought to report the late multimodality imaging and clinical outcomes of the novel poly-l-lactic-acid-based DESolve novolimus-eluting bioresorbable coronary scaffold for the treatment of de novo coronary lesions. BACKGROUND Bioresorbable scaffolds are an alternative to drug-eluting metallic stents and provide temporary vascular scaffolding, which potentially may allow ve...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 37  شماره 

صفحات  -

تاریخ انتشار 2016