Bioelectric analyses of an osseointegrated intelligent implant design system for amputees.

نویسندگان

  • Brad M Isaacson
  • Jeroen G Stinstra
  • Rob S MacLeod
  • Joseph B Webster
  • James P Beck
  • Roy D Bloebaum
چکیده

The projected number of American amputees is expected to rise to 3.6 million by 2050. Many of these individuals depend on artificial limbs to perform routine activities, but prosthetic suspensions using traditional socket technology can prove to be cumbersome and uncomfortable for a person with limb loss. Moreover, for those with high proximal amputations, limited residual limb length may prevent exoprosthesis attachment all together. Osseointegrated implant technology is a novel operative procedure which allows firm skeletal attachment between the host bone and an implant. Preliminary results in European amputees with osseointegrated implants have shown improved clinical outcomes by allowing direct transfer of loads to the bone-implant interface. Despite the apparent advantages of osseointegration over socket technology, the current rehabilitation procedures require long periods of restrictive load bearing prior which may be reduced with expedited skeletal attachment via electrical stimulation. The goal of the osseointegrated intelligent implant design (OIID) system is to make the implant part of an electrical system to accelerate skeletal attachment and help prevent periprosthetic infection. To determine optimal electrode size and placement, we initiated proof of concept with computational modeling of the electric fields and current densities that arise during electrical stimulation of amputee residual limbs. In order to provide insure patient safety, subjects with retrospective computed tomography scans were selected and three dimensional reconstructions were created using customized software programs to ensure anatomical accuracy (Seg3D and SCIRun) in an IRB and HIPAA approved study. These software packages supported the development of patient specific models and allowed for interactive manipulation of electrode position and size. Preliminary results indicate that electric fields and current densities can be generated at the implant interface to achieve the homogenous electric field distributions required to induce osteoblast migration, enhance skeletal fixation and may help prevent periprosthetic infections. Based on the electrode configurations experimented with in the model, an external two band configuration will be advocated in the future.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Osseointegrated total hip replacement connected to a lower limb prosthesis: a proof-of-concept study with three cases.

BACKGROUND Osseointegrated implants are a suitable alternative for prosthetic attachment in individuals with a transfemoral amputation, who are unable to wear a socket. However, the small bone-implant contact area, reduced muscular leverage, and osteoporosis contraindicate osseointegrated implant use in transfemoral amputees with osteoporosis and a short residuum. We report on the feasibility o...

متن کامل

Implant Survival, Adverse Events, and Bone Remodeling of Osseointegrated Percutaneous Implants for Transhumeral Amputees

BACKGROUND Osseointegrated percutaneous implants provide direct anchorage of the limb prosthesis to the residual limb. These implants have been used for the rehabilitation of transhumeral amputees in Sweden since 1995 using a two-stage surgical approach with a 6-month interval between the stages, but results on implant survival, adverse events, and radiologic signs of osseointegration and adapt...

متن کامل

Development of a Single Stage Surgical Model for Percutaneous Osseointegrated Implants for Amputees

INTRODUCTION: Limb amputation is a devastating event resulting in associated loss of everyday function and a massive change in the quality of life. To date, socket technology is the standard of care to attach, or dock, the exoprosthetic device to the residual amputated limb in an attempt to impart partial restoration of use and function. However, sockets are not without shortcomings: they can o...

متن کامل

Numerical Analysis and Comparison of Femur Bone in Osseointegrated Amputees and Non Amputees during Daily Activities Bilal

Currently available implants for osseointegration in transfemoral amputees are OPRA system and ISP System which shows many advantages over conventional socket fixation. To overcome the limiting issues like periprosthetic bone losses in osseointegration, a new design was developed. The analysis of both OPRA and New design was compared with a normal femur bone during daily activities. The results...

متن کامل

Osseointegrated total knee replacement connected to a lower limb prosthesis: 4 cases.

BACKGROUND AND PURPOSE Osseointegrated implants are an alternative for prosthetic attachment in individuals with amputation who are unable to wear a socket. However, the load transmitted through the osseointegrated fixation to the residual tibia and knee joint can be unbearable for those with transtibial amputation and knee arthritis. We report on the feasibility of combining total knee replace...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of visualized experiments : JoVE

دوره 29  شماره 

صفحات  -

تاریخ انتشار 2009