A 52-year-old man was brought to the emergency department because of multiple implantable cardioverterdefibrillator (ICD) shocks. His existing dual-chamber pacemaker was upgraded from a biventricular ICD 4 years ago. The ECG in the emergency department showed sinus tachycardia with left bundle-branch block at a rate of 135 bpm. In the emergency department, the patient received 5 additional shocks without a change in rhythm. A donut magnet was placed over the ICD. Interrogation revealed evidence of a lead fracture, and tachycardia therapies were disabled. The high-voltage lead was a Medtronic Sprint Fidelis ICD lead. Admission to the hospital was recommended for definitive correction of his problem.
منابع مشابه
THE EFFECT OF NURSE-PEER-LED SUPPORT INTERVENTION ON QUALITY OF LIFE IN PATIENTS WITH AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR: A RANDOMIZED CLINICAL TRIAL
Background & Aims: Nowadays, the quality of life and how coping and treatment of patients with an implantable cardiac defibrillator has led to concern and attempts to reduce or eliminate the problems of these patients in the people who provide health care services. This study aimed to determine the effect of nurse-peer-led support intervention on quality of life in patients with an implantable ...
متن کاملImplantable cardioverter-defibrillator in a patient with dextrocardia situs inversus
Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...
متن کاملزندگی با دفیبریلاتور قابل کاشت: یک مطالعه پدیدهشناسی
Background & Aim: Implantable Cardioverter Defibrillator is the most effective treatment for cardiac dysrhythmias. Having such life-saving device is the unique experience with important changes in the patient’ life. Understanding of these experiences can lead to success in treatment and follow- up.This study aimed to understand the experiences of patients with Implantable Cardioverter Defibri...
متن کاملThe effect of education based on Orem’s self-care model on emotional reactions in patients with Implantable Cardioverter Defibrillator (ICD)
Abstract Aim. The purpose of this study was to examine the effect of education based on Orem’s self-care model on emotional reactions in people with Implantable Cardioverter Defibrillator (ICD). Background. ICD shock is unpredictable and associated with physical pain. After receiving shock, some patients find serious psychological problems such as anxiety, depression and stress. According to Or...
متن کاملHow much scrutiny and stress can the Fidelis lead withstand?
The implantable cardioverter-defibrillator lead has undergone numerous design and manufacturing changes over the last 25 years aimed at improving patient care and outcomes. It is widely accepted that implantable cardioverterdefibrillator leads, similar to pacemaker leads, have a finite life. The design of an implantable cardioverter-defibrillator lead is more complex than that of a pacemaker le...
متن کاملArrhythmia/Electrophysiology Longevity of Sprint Fidelis Implantable Cardioverter- Defibrillator Leads and Risk Factors for Failure Implications for Patient Management
Background—Sprint Fidelis (Fidelis) implantable cardioverter-defibrillator leads are prone to fractures that have caused adverse events, primarily inappropriate shocks, and a few reported deaths. More than 100 000 patients have Fidelis leads. No independent multicenter long-term performance information exists for this lead, and single-center studies suggest that certain patients are at increase...