Health worker exposure risk during inhalation sedation with sevoflurane using the (AnaConDa®) anaesthetic conserving device.
نویسندگان
چکیده
INTRODUCTION AND OBJECTIVE Occupational exposure to sevoflurane should not exceed 2 ppm. During inhalation sedation with sevoflurane using the anaesthetic conserving device (AnaConDa(®)) in the post-anaesthesia care unit, waste gases can be reduced by gas extraction systems or scavenging devices such as CONTRAfluran™. However, the efficacy of these methods has not been clearly established. To determine the safest scenario for healthcare workers during inhalation sedation with sevoflurane in the post-surgical intensive care unit. MATERIALS AND METHODS An experimental study on occupational exposure was conducted in a post-cardiothoracic care unit during March-August 2009. The measurements were performed in four post-cardiac surgery sedated adults in post-surgical intensive care unit and four nurses at the bedside, and at four points: scenario A, inhalation sedation without gas extraction system or contrafluran as a reference scenario; scenario B, applying a gas extraction system to the ventilator; scenario C, using contrafluran; and scenario 0, performing intravenous isolation sedation. Sevoflurane concentrations were measured in the nurses' breathing area during patient care, and at 1.5 and 8 m from the ventilator using diffusive passive monitor badges. RESULTS All badges corresponding to the nurses' breathing area were below 2 ppm. Levels of sevoflurane detected using prevention systems were lower than that in the control situation. Only one determination over 2 ppm was found, corresponding to the monitor placed nearest the gas outlet of the ventilator in scenario A. Trace concentrations of sevoflurane were found in scenario 0 during intravenous sedation. CONCLUSIONS Administration of sevoflurane through the AnaConDa(®) system during inhalation sedation in post-surgical intensive care units is safe for healthcare workers, but gas extraction systems or scavenging systems, such as CONTRAfluran™ should be used to reduce occupational exposure as much as possible.
منابع مشابه
Volatile anaesthetics for analgesia and sedation in patients after abdominal surgery in ICU
Methods 37 patients on prolonged mechanical ventilation (20 males, 17 females) after abdominal surgery were followed up in ICU. The sedation level was evaluated with RASS scale and auditory evoked potentials (AEP), analgesia with the analgesia/nociceptive index (ANI), based on spectral analysis of RR intervals [2], hemodynamic parameters by impedance plethysmography. Sedation was performed by A...
متن کاملVolatile sedation with sevoflurane in intensive care patients with acute stroke or subarachnoid haemorrhage using AnaConDa®: an observational study.
BACKGROUND The anaesthetic conserving device, AnaConDa(®), allows use of inhaled anaesthetics for sedation in the intensive care unit. We prospectively measured cerebral and cardiopulmonary parameters in patients with acute stroke or subarachnoid haemorrhage during a switch from i.v. to inhalative sedation. METHODS 25 patients were switched from i.v. to an indefinite period of inhaled sedatio...
متن کاملRenal and hepatic integrity in long-term sevoflurane sedation using the anesthetic conserving device: a comparison with intravenous propofol sedation in an animal model.
INTRODUCTION Critically ill patients are sedated with intravenous agents because the use of inhaled agents is limited by their potential risk of toxicity. Increasing levels of inorganic fluorides after the metabolism of these agents have been considered potentially nephrotoxic. However, hepatic involvement after prolonged administration of sevoflurane has not yet been studied. The present study...
متن کاملImpact of the anesthetic conserving device on respiratory parameters and work of breathing in critically ill patients under light sedation with sevoflurane.
BACKGROUND Sevoflurane sedation in the intensive care unit is possible with a special heat and moisture exchanger called the Anesthetic Conserving Device (ACD) (AnaConDa; Sedana Medical AB, Uppsala, Sweden). The ACD, however, may corrupt ventilatory mechanics when used during the weaning process of intensive care unit patients. The authors compared the ventilatory effects of light-sedation with...
متن کاملAnaConDa™ and Mirus™ for intensive care sedation, 24 h desflurane versus isoflurane in one patient.
INTRODUCTION With the AnaConDa™ system, inhalational sedation in the intensive care unit has become popular. The device can be used with common intensive care unit ventilators and is inserted between the Y-piece and the patient. Liquid isoflurane or sevoflurane are delivered by a syringe pump. 90 % of anesthetic exhaled by the patient is absorbed by a reflector and resupplied during the next in...