Retrospective analysis of linear accelerator output constancy checks using process control techniques

نویسندگان

  • Taweap Sanghangthum
  • Sivalee Suriyapee
  • Somyot Srisatit
  • Todd Pawlicki
چکیده

Shewhart control charts have previously been suggested as a process control tool for use in routine linear accelerator (linac) output verifications. However, a comprehensive approach to process control has not been investigated for linac output verifications. The purpose of this work is to investigate a comprehensive process control approach to linac output constancy quality assurance (QA). The RBA-3 dose constancy check was used to verify outputs of photon beams and electron beams delivered by a Varian Clinac 21EX linac. The data were collected during 2009 to 2010. Shewhart-type control charts, exponentially weighted moving average (EWMA) charts, and capability indices were applied to these processes. The Shewhart-type individuals chart (X-chart) was used and the number of data points used to calculate the control limits was varied. The parameters tested for the EWMA charts (smoothing parameter (λ) and the control limit width (L)) were λ = 0.05, L = 2.492; λ = 0.10, L = 2.703; and λ = 0.20, L = 2.860, as well as the number of points used to estimate the initial process mean and variation. Lastly, the number of in-control data points used to determine process capability (C(p)) and acceptability (C(pk)) were investigated, comparing the first in-control run to the longest in-control run of the process data. C(p) and C(pk) values greater than 1.0 were considered acceptable. The 95% confidence intervals were reported. The X-charts detected systematic errors (e.g., device setup errors). In-control run lengths on the X-charts varied from 5 to 30 output measurements (about one to seven months). EWMA charts showed in-control runs ranging from 9 to 33 output measurements (about two to eight months). The C(p) and C(pk) ratios are higher than 1.0 for all energies, except 12 and 20 MeV. However, 10 MV and 6, 9, and 16 MeV were in question when considering the 95% confidence limits. The X-chart should be calculated using 8-12 data points. For EWMA chart, using 4 data points is sufficient to calculate the initial mean and variance of the process. The EWMA limits should be calculated with λ = 0.10, L = 2.703. At least 25-30 in-control data points should be used to calculate the C(p) and C(pk) indices.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Systematic and comprehensive analysis of the dose-response characteristics of a morning quality check of a linear accelerator and an important application of accelerator performance prediction

Background: This paper aimed to analyze the output constancy of a medical linear accelerator using PTW QUICKCHECKwebline and studied the sensitivity of the PTW QUICKCHECKwebline. Materials and Methods:  The paper statistically analyzed the output doses of 6 MV and 10 MV of photons and 6 MeV, 9 MeV, and 12 MeV of electrons from a medical linear accelerator measured before the daily treatment to ...

متن کامل

Preliminary Retrospective Analysis of Daily Tomotherapy Output Constancy Checks Using Statistical Process Control

The purpose of this study was to retrospectively evaluate the results from a Helical TomoTherapy Hi-Art treatment system relating to quality controls based on daily static and dynamic output checks using statistical process control methods. Individual value X-charts, exponentially weighted moving average charts, and process capability and acceptability indices were used to monitor the treatment...

متن کامل

Electron beam quality control using an amorphous silicon EPID.

An amorphous silicon EPID has been investigated to determine whether it is capable of quality control constancy measurements for linear accelerator electron beams. The EPID grayscale response was found to be extremely linear with dose over a wide dose range and, more specifically, for exposures of 95-100 MU. Small discrepancies of up to 0.8% in linearity were found at 6 MeV (8-15 MeV showed bet...

متن کامل

Constancy checks of well‐type ionization chambers with external‐beam radiation units

Constancy checks of a well-type ionization chamber should be performed regularly as part of a quality assurance regime. The goal of this work was to test the feasibility of using a linear accelerator and an orthovoltage unit to check the constancy of a well-type chamber's response to an external radiation source. The reproducibility, linearity with dose, variation with dose-rate, and variation ...

متن کامل

Analysis of Accuracy of Interpolation Methods in Estimating the Output Factors for Square Fields in Medical Linear Accelerator

Introduction: To estimate the accuracy levels of Lagrange, Newton backward interpolation, and linear interpolation methods in estimating the output factors for square fields used in linear accelerator for 6 MV photons at various depths. Materials and Methods: Ionization measurements were carried out in radiation field analyzer in linear accelerator for 6 MV beams at the depths of 5 and 10 cm by...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 14  شماره 

صفحات  -

تاریخ انتشار 2013