Comparison of migration behavior between single and dual lag screw implants for intertrochanteric fracture fixation

نویسندگان

  • George K Kouvidis
  • Mark B Sommers
  • Peter V Giannoudis
  • Pavlos G Katonis
  • Michael Bottlang
چکیده

BACKGROUND Lag screw cut-out failure following fixation of unstable intertrochanteric fractures in osteoporotic bone remains an unsolved challenge. This study tested if resistance to cut-out failure can be improved by using a dual lag screw implant in place of a single lag screw implant. Migration behavior and cut-out resistance of a single and a dual lag screw implant were comparatively evaluated in surrogate specimens using an established laboratory model of hip screw cut-out failure. METHODS Five dual lag screw implants (Endovis, Citieffe) and five single lag screw implants (DHS, Synthes) were tested in the Hip Implant Performance Simulator (HIPS) of the Legacy Biomechanics Laboratory. This model simulated osteoporotic bone, an unstable fracture, and biaxial rocking motion representative of hip loading during normal gait. All constructs were loaded up to 20,000 cycles of 1.45 kN peak magnitude under biaxial rocking motion. The migration kinematics was continuously monitored with 6-degrees of freedom motion tracking system and the number of cycles to implant cut-out was recorded. RESULTS The dual lag screw implant exhibited significantly less migration and sustained more loading cycles in comparison to the DHS single lag screw. All DHS constructs failed before 20,000 cycles, on average at 6,638 +/- 2,837 cycles either by cut-out or permanent screw bending. At failure, DHS constructs exhibited 10.8 +/- 2.3 degrees varus collapse and 15.5 +/- 9.5 degrees rotation around the lag screw axis. Four out of five dual screws constructs sustained 20,000 loading cycles. One dual screw specimens sustained cut-out by medial migration of the distal screw after 10,054 cycles. At test end, varus collapse and neck rotation in dual screws implants advanced to 3.7 +/- 1.7 degrees and 1.6 +/- 1.0 degrees , respectively. CONCLUSION The single and double lag screw implants demonstrated a significantly different migration resistance in surrogate specimens under gait loading simulation with the HIPS model. In this model, the double screw construct provided significantly greater resistance against varus collapse and neck rotation in comparison to a standard DHS lag screw implant.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Measurement of Tip Apex Distance and Migration of Lag Screws and Novel Blade Screw Used for the Fixation of Intertrochanteric Fractures

Fixation with a dynamic hip screw (DHS) is one of the most common methods for stabilizing intertrochanteric fractures, except for unstable and reverse oblique fracture types. However, failure is often observed in osteoporotic patients whereby the lag screw effectively 'cuts out' through the weak bone. Novel anti-migration blades have been developed to be used in combination with a lag screw ('B...

متن کامل

Comparison of Intertrochanteric Fracture Fixation with Dynamic Hip Screw and Bipolar Hemiarthroplasty Techniques

Background: At present, there is no consensus on an appropriate treatment modality for intertrochanteric fractures in the elderly with background diseases. The aim of the present study was to compare treatment outcomes of intertrochanteric fractures reduced with dynamic hip screws (DHS) and bipolar hemiarthroplasty in elderly patients with background medical conditions. Methods: In this randomi...

متن کامل

بررسی کاربرد Short barrel DHS در درمان شکستگی انترتروکانتریک

Cut out or extrusion of the lag screw from the superior aspect of head and neck of the femur is one of the most common and devastating complications of the surgery of the intertrochanteric fractures with DHS. The exact cause of this complications is unknown, but it seems to be related to osteopenia, inappropriate position of lag screw inside head of the femur and inability of DHS to slide insid...

متن کامل

Frequency and Causes of Failure of Dynamic Hip Screw Fixation for Interochanteric Fracture

Background: Hip fractures are a leading cause of death and disability among elderly. Approximately half of these injuries are intertrochanteric fractures and the incidence is continuously increasing. It is estimated that the incidence of intertrochanteric fractures will be doubled by the year 2040. Different types of implants were tried at different times for internal fixation of these fracture...

متن کامل

Dual lag screw cephalomedullary nail versus the classic sliding hip screw for the stabilization of intertrochanteric fractures. A prospective randomized study

This study is a randomized prospective study comparing two fracture fixation implants, the extramedullary sliding hip screw (SHS) and the dual lag screw cephalomedullary nail, in the treatment of intertrochanteric femoral fractures in the elderly. One hundred and sixty-five patients with low-energy intertrochanteric fractures, classified as AO/OTA 31A, were prospectively included during a 2-yea...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of Orthopaedic Surgery and Research

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2009