Fully automated quantification of hepatitis C virus (HCV) RNA in human plasma and human serum by the COBAS AmpliPrep/COBAS TaqMan system.

نویسندگان

  • Dorothea Sizmann
  • Claudia Boeck
  • Juergen Boelter
  • Daniela Fischer
  • Marion Miethke
  • Stefan Nicolaus
  • Markus Zadak
  • Reiner Babiel
چکیده

BACKGROUND HCV RNA is commonly recognized as key parameter for reliable diagnosis and treatment monitoring of HCV infection. Determination of blood HCV RNA concentrations reduces the pre-seroconversion period in the diagnosis of HCV infection and supports management of interferon alpha-based therapies of chronic HCV infection. OBJECTIVES AND STUDY DESIGN The COBAS AmpliPrep/COBAS TaqMan HCV Test combines automated extraction of nucleic acids on the COBAS AmpliPrep Instrument with real-time PCR on the COBAS TaqMan Analyzer, thus greatly reducing hands-on time during sample preparation and amplification/detection. The test, which is calibrated to the 1st International HCV WHO Standard, was evaluated for sensitivity, dynamic range, precision, matrix equivalence, genotype inclusivity, interfering substances, diagnostic and analytical specificity, as well as for correlation with two other commercial tests for HCV RNA quantification. RESULTS The COBAS AmpliPrep/COBAS TaqMan HCV Test demonstrated a >6-log dynamic range of 43-6.90 E+7 IU/mL, a sensitivity (95% hit rate) of at least 15 IU/mL for HCV WHO Standard and a comparable quantification of genotypes 1-6. HCV quantification results were in good correlation with those obtained by the COBAS AMPLICOR HCV MONITOR Test v2.0 and the VERSANT HCV RNA 3.0 test. CONCLUSIONS The fully automated COBAS AmpliPrep/COBAS TaqMan HCV Test excellently accomplishes the requirements for highly sensitive detection and reliable quantification of HCV in clinical samples and thus improves therapy monitoring and management of HCV infection.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Broad differences between the COBAS ampliprep total nucleic acid isolation-COBAS TaqMan 48 hepatitis C virus (HCV) and COBAS HCV monitor v2.0 assays for quantification of serum HCV RNA of non-1 genotypes.

Hepatitis C virus (HCV) RNA quantification is extensively used for therapeutic management of HCV chronic infection (1). The COBAS HCV Monitor v2.0 assay (HCM) was until recently one of the most used quantification methods worldwide. Two of its drawbacks are the labor-intensive manual extraction of HCV RNA and the frequent requirement of serum dilution for accurate quantification ( 5.7 log10 IU/...

متن کامل

Performance evaluation of the new Roche cobas AmpliPrep/cobas TaqMan HCV test, version 2.0, for detection and quantification of hepatitis C virus RNA.

To evaluate the analytical performance and explore the clinical applicability of the new Roche cobas AmpliPrep/cobas TaqMan HCV test, v2.0 (CAP/CTM v2.0), a platform comparison was performed on panels and diagnostic samples with the Roche cobas AmpliPrep/cobas TaqMan HCV test (CAP/CTM v1.0), the Siemens Versant HCV RNA 3.0 branched DNA (bDNA) test, the Abbott m2000 RealTime HCV assay (Realtime ...

متن کامل

Pegylated Interferon and Ribavirin in Sustained Virological Response to TaqMan HCV Assays for Prediction of and Roche Cobas AmpliPrep/Cobas Abbott RealTime Hepatitis C Virus (HCV)

Published Ahead of Print 17 December 2008. 10.1128/JCM.01753-08. 2009, 47(2):385. DOI: J. Clin. Microbiol. Masashi Mizokami Fuminaka Sugauchi, Shunsuke Nojiri, Takashi Joh and Tomoyoshi Ohno, Hiroshi Tokuda, Fuat Kurbanov, Kentaro Matsuura, Yasuhito Tanaka, Izumi Hasegawa, Chronic Hepatitis C Patients Pegylated Interferon and Ribavirin in Sustained Virological Response to TaqMan HCV Assays for ...

متن کامل

Evaluation of an automated sample preparation protocol for quantitative detection of hepatitis C virus RNA.

The COBAS AMPLIPREP instrument for automated sample preparation has recently been introduced. In this study, the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test, which includes this new molecular device, was evaluated and compared to the COBAS AMPLICOR HCV MONITOR test, which includes a manual extraction protocol. Interassay and intra-assay variation, precision, and linearity were determined, a...

متن کامل

Evaluation of the COBAS TaqMan HCV test with automated sample processing using the MagNA pure LC instrument.

The COBAS TaqMan HCV Test (TaqMan HCV; Roche Molecular Systems Inc., Branchburg, N.J.) for hepatitis C virus (HCV) performed on the COBAS TaqMan 48 Analyzer (Roche Molecular Systems) currently relies on a manual sample processing method. Implementation of an automated sample processing method would facilitate the clinical use of this test. In this study, we evaluated the performance characteris...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology

دوره 38 4  شماره 

صفحات  -

تاریخ انتشار 2007