The purpose of this study was comparing dose-volume histogram (DVH)-based plan verification methods for volumetric modulated arc therapy (VMAT) pretreatment QA. We evaluated two 3D dose reconstruction systems: ArcCHECK-3DVH system (Sun Nuclear corp.) and Varian dynalog-based dose reconstruction (DBDR) system, developed in-house. Fifteen prostate cancer patients (67.6 Gy/26 Fr), four head and neck cancer patient (66 Gy/33 Fr), and four esophagus cancer patients (60 Gy/30 Fr) treated with VMAT were studied. First, ArcCHECK measurement was performed on all plans; simultaneously, the Varian dynalog data sets that contained the actual delivered parameters (leaf positions, gantry angles, and cumulative MUs) were acquired from the Linac control system. Thereafter, the delivered 3D patient dose was reconstructed by 3DVH software (two different calculating modes were used: High Sensitivity (3DVH-HS) and Normal Sensitivity (3DVH-NS)) and in-house DBDR system. We evaluated the differences between the TPS-calculated dose and the reconstructed dose using 3D gamma passing rates and DVH dose index analysis. The average 3D gamma passing rates (3%/3 mm) between the TPS-calculated dose and the reconstructed dose were 99.1 ± 0.6%, 99.7 ± 0.3%, and 100.0 ± 0.1% for 3DVH-HS, 3DVH-NS, and DBDR, respectively. For the prostate cases, the average differences between the TPS-calculated dose and reconstructed dose in the PTV mean dose were 1.52 ± 0.50%, -0.14 ± 0.55%, and -0.03 ± 0.07% for 3DVH-HS, 3DVH-NS, and DBDR, respectively. For the head and neck and esophagus cases, the dose difference to the TPS-calculated dose caused by an effect of heterogeneity was more apparent under the 3DVH dose reconstruction than the DBDR. Although with some residual dose reconstruction errors, these dose reconstruction methods can be clinically used as effective tools for DVH-based QA for VMAT delivery.
منابع مشابه
Comparing measurement‐derived (3DVH) and machine log file‐derived dose reconstruction methods for VMAT QA in patient geometries
The purpose of this study was to compare the measurement-derived (3DVH) dose reconstruction method with machine log file-derived dose reconstruction method in patient geometries for VMAT delivery. A total of ten patient plans were selected from a regular fractionation plan to complex SBRT plans. Treatment sites in the lung and abdomen were chosen to explore the effects of tissue heterogeneity o...
متن کاملDosimetric Verification by Using the ArcCHECK System and 3DVH Software for Various Target Sizes
OBJECTIVE To investigate the usefulness of the 3DVH software with an ArcCHECK 3D diode array detector in newly designed plans with various target sizes. METHODS The isocenter dose was measured with an ion-chamber and was compared with the planned and 3DVH predicted doses. The 2D gamma passing rates were evaluated at the diode level by using the ArcCHECK detector. The 3D gamma passing rates fo...
متن کاملEvaluation of patient DVH‐based QA metrics for prostate VMAT: correlation between accuracy of estimated 3D patient dose and magnitude of MLC misalignment
The purpose of this study was to evaluate the accuracy of commercially available software, using patient DVH-based QA metrics, by investigating the correlation between estimated 3D patient dose and magnitude of MLC misalignments. We tested 3DVH software with an ArcCHECK. Two different calculating modes of ArcCHECK Planned Dose Perturbation (ACPDP) were used: "Normal Sensitivity" and "High Sensi...
متن کاملAccuracy of one algorithm used to modify a planned DVH with data from actual dose delivery
Detection and accurate quantification of treatment delivery errors is important in radiation therapy. This study aims to evaluate the accuracy of DVH based QA in quantifying delivery errors. Eighteen previously treated VMAT plans (prostate, H&N, and brain) were randomly chosen for this study. Conventional IMRT delivery QA was done with the ArcCHECK diode detector for error-free plans and plans ...
متن کاملPatient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy
Log file-based methods are attracting increasing interest owing to their ability to validate volumetric-modulated arc therapy outputs with high resolution in the leaf and gantry positions and in delivered dose. Cross-validation of these methods for comparison with measurement-based methods using the ionization chamber/ArcCHECK-3DVH software (version 3.2.0) under the same conditions of treatment...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 18 شماره
صفحات -
تاریخ انتشار 2017