A Comparative Experimental Study of the In-vitro Efficiency of Hypertonic Saline-Enhanced Hepatic Bipolar and Monopolar Radiofrequency Ablation
نویسندگان
چکیده
OBJECTIVE To compare the in-vitro efficiency of a hypertonic saline (HS)-enhanced bipolar radiofrequency (RF) system with monopolar RF applications by assessing the temperature profile and dimensions of RF-created coagulation necrosis in bovine liver. MATERIALS AND METHODS A total of 27 ablations were performed in explanted bovine livers. After placement of two 16-gauge open-perfused electrodes at an interelectrode distance of 3 cm, 5% HS was instilled into tissue at a rate of 1 mL/min through the electrode. Seventeen thermal ablation zones were created in the monopolar mode (groups A, B), and ten more were created using the two open-perfused electrodes in the bipolar mode (group C). RF was applied to each electrode for 5 mins (for a total of 10 mins, group A) or 10 mins (for a total of 20 mins, group B) at 50W in the sequential monopolar mode, or to both electrodes for 10 min in the bipolar mode (group C). During RF instillation, we measured tissue temperature at the midpoint between the two electrodes. The dimensions of the thermal ablation zones and changes in impedance and wattage during RFA were compared between the groups. RESULTS With open-perfusion electrodes, the mean accumulated energy output value was lower in the bipolar mode (group C: 26675+/-3047 Watt s) than in the monopolar mode (group A: 28778+/-1300 Watt s) but the difference was not statistically significant (p > 0.05). In the bipolar mode, there were impedance rises of more than 700 ohm during RF energy application, but in the monopolar modes, impedance did not changed markedly. In the bipolar mode, however, the temperature at the mid-point between the two probes was higher (85 degrees C) than in the monopolar modes (65 degrees C, 80 degrees C for group A, B, respectively) (p < 0.05). In addition, in HS-enhanced bipolar RFA (group C), the shortest diameter at the midpoint between the two electrodes was greater than in either of the monopolar modes: 5.4+/-5.6 mm (group A); 28.8+/-8.2 mm (group B); 31.2+/-7.6 mm (group C) (p < 0.05) CONCLUSION Using an open perfusion system, HS-enhanced bipolar RFA more efficiently created larger areas of thermal ablation and higher tissue temperatures than monopolar RFA.
منابع مشابه
Saline-Enhanced Hepatic Radiofrequency Ablation Using a Perfused-Cooled Electrode: Comparison of Dual Probe Bipolar Mode with Monopolar and Single Probe Bipolar Modes
OBJECTIVE To determine whether saline-enhanced dual probe bipolar radiofrequency ablation (RFA) using perfused-cooled electrodes shows better in-vitro efficiency than monopolar or single probe bipolar RFA in creating larger coagulation necrosis. MATERIALS AND METHODS RF was applied to excised bovine livers in both bipolar and monopolar modes using a 200 W generator (CC-3; Radionics) and the p...
متن کاملNeedle-based ablation of renal parenchyma using microwave, cryoablation, impedance- and temperature-based monopolar and bipolar radiofrequency, and liquid and gel chemoablation: laboratory studies and review of the literature.
BACKGROUND AND PURPOSE Small renal tumors are often serendipitously detected during the screening of patients for renal or other disease entities. Rather than perform a radical or partial nephrectomy for these diminutive lesions, several centers have begun to explore a variety of ablative energy sources that could be applied directly via a percutaneously placed needle-like probe. To evaluate th...
متن کاملHepatic Radiofrequency Ablation Using Multiple Probes: Ex Vivo and In Vivo Comparative Studies of Monopolar versus Multipolar Modes
OBJECTIVE We wanted to compare the efficiency of multipolar radiofrequency ablation (RFA) using three perfused-cooled electrodes with multiple overlapping and simultaneous monopolar techniques for creating an ablation zone in ex vivo bovine livers and in in vivo porcine livers. MATERIALS AND METHODS In the ex vivo experiments, we used a 200 W generator (Valleylab, CC-3 model) and three perfus...
متن کاملRadio Frequency Ablation in the Rabbit Lung Using Wet Electrodes: Comparison of Monopolar and Dual Bipolar Electrode Mode
OBJECTIVE To compare the effect of radio frequency ablation (RFA) on the dimensions of radio frequency coagulation necrosis in a rabbit lung using a wet electrode in monopolar mode with that in dual electrode bipolar mode at different infusion rates (15 mm/hr versus 30 ml/hr) and saline concentrations (0.9% normal versus 5.8% hypertonic saline). MATERIALS AND METHODS Fifty ablation zones (one...
متن کاملHeat Sink Effect on Tumor Ablation Characteristics as Observed in Monopolar Radiofrequency, Bipolar Radiofrequency, and Microwave, Using Ex Vivo Calf Liver Model
Thermal ablation of liver tumors near large blood vessels is affected by the cooling effect of blood flow, leading to incomplete ablation. Hence, we conducted a comparative investigation of heat sink effect in monopolar (MP) and bipolar (BP) radiofrequency ablation (RFA), and microwave (MW) ablation devices.With a perfused calf liver, the ablative performances (volume, mass, density, dimensions...