A new method and device of aligning patient setup lasers in radiation therapy

نویسندگان

  • Ui‐Jung Hwang
  • Kwanghyun Jo
  • Young Kyung Lim
  • Jung Won Kwak
  • Sang Hyoun Choi
  • Chiyoung Jeong
  • Mi Young Kim
  • Jong Hwi Jeong
  • Dongho Shin
  • Se Byeong Lee
  • Jeong‐Hoon Park
  • Sung Yong Park
  • Siyong Kim
چکیده

The aim of this study is to develop a new method to align the patient setup lasers in a radiation therapy treatment room and examine its validity and efficiency. The new laser alignment method is realized by a device composed of both a metallic base plate and a few acrylic transparent plates. Except one, every plate has either a crosshair line (CHL) or a single vertical line that is used for alignment. Two holders for radiochromic film insertion are prepared in the device to find a radiation isocenter. The right laser positions can be found optically by matching the shadows of all the CHLs in the gantry head and the device. The reproducibility, accuracy, and efficiency of laser alignment and the dependency on the position error of the light source were evaluated by comparing the means and the standard deviations of the measured laser positions. After the optical alignment of the lasers, the radiation isocenter was found by the gantry and collimator star shots, and then the lasers were translated parallel to the isocenter. In the laser position reproducibility test, the mean and standard deviation on the wall of treatment room were 32.3 ± 0.93 mm for the new method whereas they were 33.4 ± 1.49 mm for the conventional method. The mean alignment accuracy was 1.4 mm for the new method, and 2.1 mm for the conventional method on the walls. In the test of the dependency on the light source position error, the mean laser position was shifted just by a similar amount of the shift of the light source in the new method, but it was greatly magnified in the conventional method. In this study, a new laser alignment method was devised and evaluated successfully. The new method provided more accurate, more reproducible, and faster alignment of the lasers than the conventional method.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Utilization of Electronic Portal Imaging Device (EPID) For Setup Verification and Determination of Setup Margin in Head and Neck Radiation Therapy

Introduction: Radiation therapy involves a multistep procedure; therefore, the error in patient set up is an inherent part of the treatment. Main purpose of this study was to determine the clinical target volume (CTV) to planning target volume (PTV) in head and neck cancer patients. Material and Methods: A total of 15 patients who had daily p...

متن کامل

Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model

Introduction: A new method based on image registration technique and an intelligent correlation model to calculate. The present study aimed to propose inter- and intra-fraction motion errors in order to address the limitations of conventional Patient positioning methods. Material and Methods: The configuration of the markerless method wa...

متن کامل

A Simulation Study on Patient Setup Errors in External Beam Radiotherapy Using an Anthropomorphic 4D Phantom

Introduction Patient set-up optimization is required in radiotherapy to fill the accuracy gap between personalized treatment planning and uncertainties in the irradiation set-up. In this study, we aimed to develop a new method based on neural network to estimate patient geometrical setup using 4-dimensional (4D) XCAT anthropomorphic phantom. Materials and Methods To access 4D modeling of motion...

متن کامل

Design, Simulation and Manufacture of a Multileaf Collimator to Confirm the Target Volumes in Intensity Modulated Radiation Therapy

Introduction: Intensity modulated radiation therapy (IMRT) is one of the cancer treatment methods. It is important to selectively aim at the target in this way, which can be performed using a multileaf collimator (MLC). Materials and Methods: In order to specifically irradiate the target volume in radiotherapy to reduce the patient absorbed dose, the use of multileaf collimator has been investi...

متن کامل

High Intensity Lasers in Periodontics: A Review

Background and Aim: Lasers can serve as a new powerful tool in dentistry. High intensity laser therapy is becoming the treatment of choice compared with conventional periodontal therapy due to accurate incision and soft and hard tissue ablation, hemostasis, reduction of postoperative pain and infection, optimal debridement, and enhanced healing. Although high intensity laser therapy assumes to ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016