Radiofrequency-targeted vertebral augmentation versus traditional balloon kyphoplasty: radiographic and morphologic outcomes of an ex vivo biomechanical pilot study [Erratum]

نویسندگان

  • Brian E Dalton
  • Andrew C Kohm
  • Larry E Miller
  • Jon E Block
  • Robert D Poser
چکیده

PURPOSE Traditional balloon kyphoplasty (BK) is a common treatment for symptomatic vertebral compression fractures. The purpose of this study was to compare a novel vertebral augmentation technique, radiofrequency-targeted vertebral augmentation (RF-TVA), to BK for restoration of vertebral height, cavity creation, and polymethylmethacrylate (PMMA) delivery and interdigitation into the surrounding trabeculae. METHODS This ex vivo biomechanical pilot study utilized 16 osteoporotic cadaveric vertebral bodies in a standardized fracture model to compare unipedicular RF-TVA (n = 8) to bipedicular BK (n = 8). Four specimens from each group were tested in loaded and unloaded conditions. All specimens were imaged, assessed for height restoration, and sectioned to observe PMMA distribution. A subset of specimens underwent computed tomography scanning to assess cavity creation and trabecular architecture prior to cement delivery. RESULTS Anterior height restoration was greater with RF-TVA (median: 84%, interquartile range: 62%-95%) compared to BK (median: 69%, interquartile range: 60%-81%), although the difference did not achieve statistical significance (P = 0.16). Anterior height restoration was numerically greater under loaded (median: 70% versus 66%) and unloaded (median: 94% versus 77%) conditions with RF-TVA versus BK. RF-TVA produced more discrete cavities and less native trabecular destruction compared to marked trabecular destruction observed with BK. RF-TVA consistently showed a well-identified focal area of PMMA with an extensive peripheral zone of PMMA interdigitation, providing mechanical interlock into the adjacent intact trabecular matrix. In contrast, BK yielded little evidence of PMMA interdigitation beyond the boundaries created by the balloon tamp due to the crushed trabecular bone peripherally. CONCLUSION RF-TVA achieves favorable vertebral height restoration with targeted PMMA delivery and less trabecular destruction compared to BK. RF-TVA has potential clinical utility in the aging population with painful osteoporotic vertebral fracture.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison between radiofrequency targeted vertebral augmentation and balloon kyphoplasty in the treatment of vertebral compression fractures: addressing factors that affect cement extravasation and distribution.

BACKGROUND Both vertebroplasty and balloon kyphoplasty have been described for treatment of vertebral compression fractures. Vertebroplasty is known for its high leakage rate compared with balloon kyphoplasty. In vitro studies have shown that high-viscosity cements significantly decrease the incidence of cement leakage and increase the predictability of cement fill in cancellous bonelike substr...

متن کامل

Cement interdigitation and bone-cement interface after augmenting fractured vertebrae: A cadaveric study

BACKGROUND The treatment of painful osteoporotic vertebral compression fractures with transpedicular cement augmentation has grown significantly over the last 20 years. There is still uncertainty about long-term and midterm effects of polymethyl methacrylate in trabecular bone. Preservation of the trabecular structures, as well as interdigitation of the cement with the surrounding bone, therefo...

متن کامل

A biomechanical comparison of kyphoplasty using a balloon bone tamp versus an expandable polymer bone tamp in a deer spine model.

We performed a biomechanical study to compare the augmentation of isolated fractured vertebral bodies using two different bone tamps. Compression fractures were created in 21 vertebral bodies harvested from red deer after determining their initial strength and stiffness, which was then assessed after standardised bipedicular vertebral augmentation using a balloon or an expandable polymer bone t...

متن کامل

A randomized trial comparing balloon kyphoplasty and vertebroplasty for vertebral compression fractures due to osteoporosis.

BACKGROUND AND PURPOSE Several trials have compared vertebral augmentation with nonsurgical treatment for vertebral compression fractures. This trial compares the efficacy and safety of balloon kyphoplasty and vertebroplasty. MATERIALS AND METHODS Patients with osteoporosis with 1-3 acute fractures (T5-L5) were randomized and treated with kyphoplasty (n = 191) or vertebroplasty (n = 190) and ...

متن کامل

An ex vivo biomechanical evaluation of an inflatable bone tamp used in the treatment of compression fracture.

STUDY DESIGN Ex vivo biomechanical study using osteoporotic cadaveric vertebral bodies. OBJECTIVES To determine if the inflatable bone tamp (tamp) restores height to compressed vertebral bodies and to compare the biomechanical properties of isolated, fractured osteoporotic vertebral bodies treated by kyphoplasty (tamp) or vertebroplasty. SUMMARY OF BACKGROUND DATA Previous biomechanical stu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2012