Improved separation of the 209 polychlorinated biphenyl congeners using comprehensive two-dimensional gas chromatography-time-of-flight mass spectrometry.
نویسندگان
چکیده
The separation of the 209 polychlorinated biphenyl (PCB) congeners has been studied using comprehensive two-dimensional gas chromatography coupled to time-of-flight mass spectrometry (GC x GC-TOFMS). Four column combinations based on thermally stable phases, DB-1/HT-8, DB-XLB/HT-8, DB-XLB/BPX-50, and HT-8/BPX-50, have been investigated. The HT-8/BPX-50 set produced the best separation. The distribution of the 100 to 150 ms wide peaks was highly structured in the chromatographic space and based on the degree of ortho-substitution within each separated homologue series. A total of 192 congeners were resolved in 146 min (1.3 analyte per min) using this column set. Eight coelutions involved 17 congeners. Among them, seven congeners were present in Aroclors at levels >1.0 wt.% (CBs 33, 47, 48, 95, 97, 163, 187). Except for CBs 47 and 48, none of the major constituents of commercial mixtures were coeluting. CB 138 was well separated from CBs 163 and 164 in the second dimension. For all column sets, CBs 20, 33, and 109 always coeluted with other PCBs. The 12 toxic dioxin-like congeners (CBs 77, 81, 105, 114, 118, 123, 126, 156, 157, 167, 169, 189), and the seven European Union marker PCBs (CBs 28, 52, 101, 118, 138, 153, 180) were separated from any interfering congeners. This was not the case for the other investigated column sets.
منابع مشابه
Fingerprinting polychlorinated biphenyls in environmental samples using comprehensive two-dimensional gas chromatography with time-of-flight mass spectrometry.
A GC × GC-TOFMS installed with a Rtx-PCB (60 m × 0.18 mm × 0.18 μm) in the first dimension and Rxi- 17 (1.5 m × 0.1 mm × 0.1 μm) column in the second dimension was used to separate 188 out of 209 congeners. A further 12 congeners were identified through additional data processing resulting in the identification of a total of 200 congeners. However, caution is advised if these 12 congeners were ...
متن کاملPolychlorinated biphenyls (PCBs) analysis using a miniaturized high-resolution time-of-flight mass spectrometer "MULTUM-S II".
In this study, we conducted polychlorinated biphenyls (PCBs) analysis using fast gas chromatography (GC)/high-resolution mass spectrometry (HRMS). Mass spectrometry (MS) was performed with a miniature multi-turn time-of-flight (TOF) analyzer called "MULTUM-S II". MULTUM-S II is truly a portable high resolution mass spectrometer. The mass spectrometer's high resolution capability is due to its t...
متن کاملIdentifying the provenance of Leach's storm petrels in the North Atlantic using polychlorinated biphenyl signatures derived from comprehensive two-dimensional gas chromatography with time-of-flight mass spectrometry.
PCB signatures can be used for source identification, exposure studies, age dating and bio-monitoring. This study uses comprehensive two-dimensional gas chromatography with time-of-flight mass spectrometry (GCxGC-ToFMS) to produce a PCB signature comprised of over 80 PCBs for individual Leach's storm petrels (Oceanodroma leucorhoa). The Leach's storm petrel is a relatively small, elusive, under...
متن کاملFingerprinting polychlorinated biphenyls (PCBs) in environmental samples using comprehensive two-dimensional gas chromatography with time-of- flight mass spectrometry
A GCxGC-TOFMS installed with a Rtx-PCB (60 m x 0.18 mm x 0.18 μm) in the first dimension and Rxi-17 (1.5 m x 0.1 mm x 0.1 μm) column in the second dimension was used to separate 188 out of 209 congeners. A further 12 congeners were identified through additional data processing resulting in the identification of a total of 200 congeners. However, caution is advised if these 12 congeners were to ...
متن کاملPotential application of gas chromatography/tandem mass spectrometry in the measurement of coeluting isomers.
Despite the unprecedented popularity of separation chromatography, the measurement of coeluting isomeric chemicals remains an extremely difficult task. We developed an analytical scheme capable of measuring two coeluting isomers using a single chromatographic column and a gas chromatography/tandem mass spectrometry system. The protocol utilized two product ion fragments generated from a common ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of chromatography. A
دوره 1040 2 شماره
صفحات -
تاریخ انتشار 2004