Heat Sink Effect on Tumor Ablation Characteristics as Observed in Monopolar Radiofrequency, Bipolar Radiofrequency, and Microwave, Using Ex Vivo Calf Liver Model

نویسندگان

  • Krishna Pillai
  • Javid Akhter
  • Terence C. Chua
  • Mena Shehata
  • Nayef Alzahrani
  • Issan Al-Alem
  • David L. Morris
  • Hui-Xiong Xu.
چکیده

Thermal ablation of liver tumors near large blood vessels is affected by the cooling effect of blood flow, leading to incomplete ablation. Hence, we conducted a comparative investigation of heat sink effect in monopolar (MP) and bipolar (BP) radiofrequency ablation (RFA), and microwave (MW) ablation devices.With a perfused calf liver, the ablative performances (volume, mass, density, dimensions), with and without heat sink, were measured. Heat sink was present when the ablative tip of the probes were 8.0 mm close to a major hepatic vein and absent when >30 mm away. Temperatures (T1 and T2) on either side of the hepatic vein near the tip of the probes, heating probe temperature (T3), outlet perfusate temperature (T4), and ablation time were monitored.With or without heat sink, BP radiofrequency ablated a larger volume and mass, compared with MP RFA or MW ablation, with latter device producing the highest density of tissue ablated. MW ablation produced an ellipsoidal shape while radiofrequency devices produced spheres.Percentage heat sink effect in Bipolar radiofrequency : Mono-polar radiofrequency : Microwave was (Volume) 33:41:22; (mass) 23:56:34; (density) 9.0:26:18; and (relative elipscity) 5.8:12.9:1.3, indicating that BP and MW devices were less affected.Percentage heat sink effect on time (minutes) to reach maximum temperature (W) = 13.28:9.2:29.8; time at maximum temperature (X) is 87:66:16.66; temperature difference (Y) between the thermal probes (T3) and the temperature (T1 + T2)/2 on either side of the hepatic vessel was 100:87:20; and temperature difference between the (T1 + T2)/2 and temperature of outlet circulating solution (T4), Z was 20.33:30.23:37.5.MW and BP radiofrequencies were less affected by heat sink while MP RFA was the most affected. With a single ablation, BP radiofrequency ablated a larger volume and mass regardless of heat sink.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

No-Touch Radiofrequency Ablation: A Comparison of Switching Bipolar and Switching Monopolar Ablation in Ex Vivo Bovine Liver

OBJECTIVE To evaluate the feasibility, efficiency, and safety of no-touch switching bipolar (SB) and switching monopolar (SM) radiofrequency ablation (RFA) using ex vivo bovine livers. MATERIALS AND METHODS A pork loin cube was inserted as a tumor mimicker in the bovine liver block; RFA was performed using the no-touch technique in the SM (group A1; 10 minutes, n = 10, group A2; 15 minutes, n...

متن کامل

Thermographic real-time-monitoring of surgical radiofrequency and microwave ablation in a perfused porcine liver model

Radiofrequency ablation (RFA) and microwave ablation (MWA) are currently the dominant modalities to treat unresectable liver tumors. Monitoring the ablation process with b-mode-sonography is often hampered by artefacts. Furthermore, vessels may cause cooling in the adjacent tumor target (heat-sink-effect) with risk of local recurrence. The present study evaluated infrared-thermography to monito...

متن کامل

Bipolar radiofrequency ablation of the kidney: comparison with monopolar radiofrequency ablation.

PURPOSE We report initial ex vivo and in vivo studies using bipolar radiofrequency (RF) ablation of porcine kidneys. An internal ground electrode is positioned in the kidney opposite the RF electrode, resulting in ablation of all the intervening renal tissue. MATERIALS AND METHODS Ex vivo preparations of 10 porcine kidneys were perfused continuously with Ringer's solution and treated with eit...

متن کامل

Ex vivo evaluation of a bipolar application concept for radiofrequency ablation.

BACKGROUND Bipolar radiofrequency ablation (RFA) can avoid complications such as thermal tissue damage, a possible consequence of monopolar RFA. However, basic studies about the dosage/effect relationship of bipolar systems are missing. This is systematic research on ex vivo beef livers to find which capacity parameters produce high energy in the tissue and achieve large lesion volumes. MATER...

متن کامل

Software Tool for the Analysis of the Coagulation Zone from Multipolar Radiofrequency Ablation

Percutaneous radiofrequency (RF) ablation is becoming a standard minimal invasive procedure for treatment of primary liver tumors. In clinical practice, knowledge about geometry and expected size of the coagulation is essential to prevent local tumor recurrence. Particularly the heat-sink effects of large vessels in the vicinity of the electrode may lead to a smaller size of the coagulation zon...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 94  شماره 

صفحات  -

تاریخ انتشار 2015