What does built-in software of home ventilators tell us? An observational study of 150 patients on home ventilation.
نویسندگان
چکیده
BACKGROUND Recent home ventilators are equipped with built-in software which provides data such as compliance, estimations of leaks, tidal volume, minute ventilation, respiratory rate, apnea and apnea-hypopnea indexes, and percentage of inspirations triggered by the patient (or ventilator). However, for many of these variables, there is neither consensus nor documentation as to what is to be expected in a population of stable patients under noninvasive ventilation (NIV). OBJECTIVES To document the values and distribution of specific items downloaded from ventilator monitoring software, by diagnostic category. METHODS Analysis of data downloaded from home ventilators in clinically stable patients under long-term NIV, during elective home visits by specialized nurses. RESULTS Data were collected from home ventilators of 150 patients with chronic obstructive pulmonary disease (n = 32), overlap syndrome (n = 29), obesity-hypoventilation (n = 38), neuromuscular disorders (n = 19), restrictive disorders (n = 21), and central sleep apnea syndrome (n = 11). On average, leaks were low, being lowest in patients with facial masks (vs. nasal masks), and increased with older age. Compliance was excellent in all groups. Patients with neuromuscular diseases triggered their ventilators less and tended to be 'captured', while other groups triggered at least half of inspiratory cycles. Most patients had a respiratory rate just slightly above the back-up rate. Residual apneas and hypopneas were highest in patients with central apneas. CONCLUSIONS Built-in software of home ventilators provides the clinician with new parameters, some of which are a useful adjunct to recommended tools for monitoring NIV and may contribute to a better understanding of residual hypoventilation and/or desaturations. However, an independent validation of the accuracy of this information is mandatory.
منابع مشابه
Home mechanical ventilation through mask: monitoring leakage and nocturnal oxygenation at home.
BACKGROUND Leakage is common in patients receiving home mechanical ventilation (HMV) via a face mask. Although pressure ventilators have partial compensatory capacity, excessive leakage can compromise the effectiveness of treatment. Home ventilators are equipped with built-in software which provides information on leakage. However, the values of leakage and their effects in routine clinical pra...
متن کاملHome Mechanical Ventilation through Mask: Monitoring Leakage and Nocturnal Oxygenation at Home
Background: Leakage is common in patients receiving home mechanical ventilation (HMV) via a face mask. Although pressure ventilators have partial compensatory capacity, excessive leakage can compromise the effectiveness of treatment. Home ventilators are equipped with built-in software which provides information on leakage. However, the values of leakage and their effects in routine clinical pr...
متن کاملEvaluation of ventilators for mouthpiece ventilation in neuromuscular disease.
BACKGROUND Daytime mouthpiece ventilation is a useful adjunct to nocturnal noninvasive ventilation (NIV) in patients with neuromuscular disease. The aims of the study were to analyze the practice of mouthpiece ventilation and to evaluate the performance of ventilators for mouthpiece ventilation. METHODS Practice of mouthpiece ventilation was assessed by a questionnaire, and the performance of...
متن کاملEffect of leak and breathing pattern on the accuracy of tidal volume estimation by commercial home ventilators: a bench study.
BACKGROUND New home ventilators are able to provide clinicians data of interest through built-in software. Monitoring of tidal volume (VT) is a key point in the assessment of the efficacy of home mechanical ventilation. OBJECTIVE To assess the reliability of the VT provided by 5 ventilators in a bench test. METHODS Five commercial ventilators from 4 different manufacturers were tested in pr...
متن کاملASSISTED VENTILATION Performance of mechanical ventilators at the patient’s home: a multicentre quality control study
Background: Quality control procedures vary considerably among the providers of equipment for home mechanical ventilation (HMV). Methods: A multicentre quality control survey of HMV was performed at the home of 300 patients included in the HMV programmes of four hospitals in Barcelona. It consisted of three steps: (1) the prescribed ventilation settings, the actual settings in the ventilator co...