Crimping and deployment of balloon-expandable valved stents are responsible for the increase in the hydraulic conductance of leaflets.
نویسندگان
چکیده
OBJECTIVES Leaflet injury has been documented to occur during the deployment of valved stents (VSs). The pathological aspects, however, of this injury are difficult to quantify. Conversely, the hydraulic conductance of a (pericardial) membrane may be easily determined. The impact of crimping and deployment of VS on this parameter was therefore investigated. METHODS Bovine pericardial square (25 × 25 mm) patches were placed within a pressure chamber and their hydraulic conductance was determined. The influence of the pressure gradient and tissue thickness on this parameter was analysed. Six balloon-expandable VS were constructed. The hydraulic conductance of their bovine pericardial leaflets was determined before and after VS crimping and deployment in four of them. Pericardial leaflets of non-crimped VS were used as controls. RESULTS Hydraulic conductance increased insignificantly with the pressure level within the chamber: from 128 ± 26.9 ml/h/m(2)/mmHg at a pressure of 50 mmHg to 232.3 ± 51.9 ml/h/m(2)/mmHg at a pressure of 250 mmHg (P = 0.117). Hydraulic conductance was not correlated to pericardial thickness, for thickness measurements ranging from 0.34 to 0.76 mm. The hydraulic conductance of VS leaflets significantly increased immediately after crimping from 45.2 ± 7.6 to 667.9.0 ± 527.2 ml/h/m(2)/mmHg (P < 0.001). This increase was still observed 24 h after VS deployment. No change in hydraulic conductance occurred in the control group. CONCLUSIONS The determination of the hydraulic conductance of pericardial patches was easy to perform, reproducible and not influenced by tissue thickness. The hydraulic conductance of pericardial leaflets dramatically increased after VS crimping and deployment. This parameter might be, in the future, a useful noninvasive tool in studying leaflet trauma.
منابع مشابه
Traumatic leaflet injury during the use of percutaneous valves: a comparative study of balloon- and self-expandable valved stents.
OBJECTIVES No comparison of balloon- or self-expandable valved stents (VSs) regarding tissue injury (if any) has been reported yet. The objective was to evaluate the occurrence and compare the severity of traumatic injury to leaflets from balloon- or self-expandable VSs. METHODS Twelve homemade VSs were used for this experiment. These three-leaflet bovine pericardial bioprostheses had either ...
متن کاملNovel self-expandable, stent-based transcatheter pulmonic valve: a preclinical animal study.
BACKGROUND Because transcatheter implantation of pulmonary valve is indicated for limited-size dysfunctional right ventricular outflow tract only as a balloon-expandable stent, we investigated the feasibility of a large-diameter self-expandable valved stent and the durability of the valve after >6 months. METHODS We made a nitinol-wire-based, self-expandable valved stent with leaflets made fr...
متن کاملOn the importance of modeling stent procedure for predicting arterial mechanics.
The stent-artery interactions have been increasingly studied using the finite element method for better understanding of the biomechanical environment changes on the artery and its implications. However, the deployment of balloon-expandable stents was generally simplified without considering the balloon-stent interactions, the initial crimping process of the stent, its overexpansion routinely u...
متن کاملBalloon-expanding stent and delivery system for transcatheter aortic valve implantation: An animal study
OBJECTIVE To evaluate the feasibility and satefy of transcatheter aortic valve implantation in animals by using a new balloon-expanding valved stent. METHODS The balloon-expandable stent is made from cobalt-based alloy material and designed with a tubular, slotted structure. Fresh bovine pericardium was treated, sutured and fixed on the balloon-expandable stent. Ten healthy sheep (five males ...
متن کاملPatient-Specific Numerical Model of Calcific Aortic Stenosis and Its Treatment by Balloon-Expandable Transcatheter Aortic Valve: Effect of Positioning on the Anchorage
Despite the promising outcomes of transcatheter aortic valve replacement (TAVR), adverse events may occur as a result of suboptimal placement. The aim of this study is to evaluate the effect of various TAVR deployment positions on the risk for intra-procedural migration and of post-procedural paravalvular regurgitation. Finite Element (FE) and Computational Fluid Dynamics (CFD) models are prese...