Comparison of Matrix Frequency-Doubling Technology (FDT) Perimetry with the SWEDISH Interactive Thresholding Algorithm (SITA) Standard Automated Perimetry (SAP) in Mild Glaucoma
نویسندگان
چکیده
This study aimed to compare second-generation frequency-doubling technology (FDT) perimetry with standard automated perimetry (SAP) in mild glaucoma. Forty-seven eyes of 47 participants who had mild visual field defect by SAP were included in this study. All participants were examined using SITA 24-2 (SITA-SAP) and matrix 24-2 (Matrix-FDT). The correlations of global indices and the number of defects on pattern deviation (PD) plots were determined. Agreement between two sets regarding the stage of visual field damage was assessed. Pearson's correlation, intra-cluster comparison, paired t-test, and 95% limit of agreement were calculated. Although there was no significant difference between global indices, the agreement between the two devices regarding the global indices was weak (the limit of agreement for mean deviation was -6.08 to 6.08 and that for pattern standard deviation was -4.42 to 3.42). The agreement between SITA-SAP and Matrix-FDT regarding the Glaucoma Hemifield Test (GHT) and the number of defective points in each quadrant and staging of the visual field damage was also weak. Because the correlation between SITA-SAP and Matrix-FDT regarding global indices, GHT, number of defective points, and stage of the visual field damage in mild glaucoma is weak, Matrix-FDT cannot be used interchangeably with SITA-SAP in the early stages of glaucoma.
منابع مشابه
Comparison of standard automated perimetry, frequency-doubling technology perimetry, and short-wavelength automated perimetry for detection of glaucoma.
PURPOSE To compare the performance of standard automated perimetry (SAP), frequency-doubling technology (FDT) perimetry, and short-wavelength automated perimetry (SWAP) in detecting glaucoma. METHODS One hundred thirty-two eyes of 95 glaucoma patients and 37 normal subjects had retinal nerve fiber layer (RNFL) imaging and visual field testing by SAP, Matrix FDT perimetry, and Swedish interact...
متن کاملComparison of visual field test results obtained through Humphrey matrix frequency doubling technology perimetry versus standard automated perimetry in healthy children
AIMS The aim of this study was to compare the visual field test results in healthy children obtained via the Humphrey matrix 24-2 threshold program and standard automated perimetry (SAP) using the Swedish interactive threshold algorithm (SITA)-Standard 24-2 test. MATERIALS AND METHODS This prospective study included 55 healthy children without ocular or systemic disorders who underwent both S...
متن کاملEXTENDED REPORT Clinical evaluation of frequency doubling technology perimetry using the Humphrey Matrix 24-2 threshold strategy
Aims: To evaluate performance of frequency doubling technology (FDT) perimetry using the Humphrey Matrix 24-2 thresholding program in a hospital eye service (HES) glaucoma clinic. Methods: A random sample of individuals referred consecutively to the HES for suspected glaucoma were examined with 24-2 threshold FDT in addition to routine clinical tests. The discriminatory power of FDT and standar...
متن کاملDetecting visual function abnormalities using the Swedish interactive threshold algorithm and matrix perimetry in eyes with glaucomatous appearance of the optic disc.
OBJECTIVE To compare the ability of 24-2 frequency-doubling perimetry (FDP-Matrix) with standard automated perimetry with the Swedish interactive threshold algorithm (SAP-SITA) in detection of visual function abnormalities in patients with glaucomatous-appearing optic discs (GAOD). METHODS This observational case-control study included 80 patients with GAOD and 54 control subjects diagnosed b...
متن کاملClinical evaluation of frequency doubling technology perimetry using the Humphrey Matrix 24-2 threshold strategy.
AIMS To evaluate performance of frequency doubling technology (FDT) perimetry using the Humphrey Matrix 24-2 thresholding program in a hospital eye service (HES) glaucoma clinic. METHODS A random sample of individuals referred consecutively to the HES for suspected glaucoma were examined with 24-2 threshold FDT in addition to routine clinical tests. The discriminatory power of FDT and standar...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 6 شماره
صفحات -
تاریخ انتشار 2017