In vivo 1H NMR spectroscopy of the human brain at high magnetic fields: metabolite quantification at 4T vs. 7T.

نویسندگان

  • Ivan Tkác
  • Gülin Oz
  • Gregor Adriany
  • Kamil Uğurbil
  • Rolf Gruetter
چکیده

A comprehensive comparative study of metabolite quantification from the human brain was performed on the same 10 subjects at 4T and 7T using MR scanners with identical consoles, the same type of RF coils, and identical pulse sequences and data analysis. Signal-to-noise ratio (SNR) was increased by a factor of 2 at 7T relative to 4T in a volume of interest selected in the occipital cortex using half-volume quadrature radio frequency (RF) coils. Spectral linewidth was increased by 50% at 7T, which resulted in a 14% increase in spectral resolution at 7T relative to 4T. Seventeen brain metabolites were reliably quantified at both field strengths. Metabolite quantification at 7T was less sensitive to reduced SNR than at 4T. The precision of metabolite quantification and detectability of weakly represented metabolites were substantially increased at 7T relative to 4T. Because of the increased spectral resolution at 7T, only one-half of the SNR of a 4T spectrum was required to obtain the same quantification precision. The Cramér-Rao lower bounds (CRLB), a measure of quantification precision, of several metabolites were lower at both field strengths than the intersubject variation in metabolite concentrations, which resulted in a strong correlation between metabolite concentrations of individual subjects measured at 4T and 7T.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

T1 and T2 Metabolite Relaxation Times in Normal Brain at 3T and 7T

Proton magnetic resonance spectroscopy (MRS) is a powerful noninvasive tool that has been widely used for the assessment of brain metabolites and the investigation of normal and abnormal metabolism in brain tissue. With the availability of improved hardware and acquisition techniques, in vivo single-voxel 1H-MRS and multivoxel magnetic resonance spectroscopic imaging (MRSI) can offer new possib...

متن کامل

Handling macromolecule signals in the quantification of the neurochemical profile.

In vivo localized proton magnetic resonance spectroscopy (1H MRS) became a powerful and unique technique to non-invasively investigate brain metabolism of rodents and humans. The main goal of 1H MRS is the reliable quantification of concentrations of metabolites (neurochemical profile) in a well-defined region of the brain. The availability of very high magnetic field strengths combined with th...

متن کامل

Investigating brain metabolism at high fields using localized 13C NMR spectroscopy without 1H decoupling.

Most in vivo 13C NMR spectroscopy studies in the brain have been performed using 1H decoupling during acquisition. Decoupling imposes significant constraints on the experimental setup (particularly for human studies at high magnetic field) in order to stay within safety limits for power deposition. We show here that incorporation of the 13C label from 13C-labeled glucose into brain amino acids ...

متن کامل

1H NMR spectroscopy of rat brain in vivo at 14.1Tesla: improvements in quantification of the neurochemical profile.

Ultra-short echo-time proton single voxel spectra of rat brain were obtained on a 14.1T 26 cm horizontal bore system. At this field, the fitted linewidth in the brain tissue of adult rats was about 11 Hz. New, separated resonances ascribed to phosphocholine, glycerophosphocholine and N-acetylaspartate were detected for the first time in vivo in the spectral range of 4.2-4.4 ppm. Moreover, impro...

متن کامل

7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images.

Signal-to-noise ratio (SNR), RF field (B(1)), and RF power requirement for human head imaging were examined at 7T and 4T magnetic field strengths. The variation in B(1) magnitude was nearly twofold higher at 7T than at 4T ( approximately 42% compared to approximately 23%). The power required for a 90 degrees pulse in the center of the head at 7T was approximately twice that at 4T. The SNR avera...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Magnetic resonance in medicine

دوره 62 4  شماره 

صفحات  -

تاریخ انتشار 2009