Heat and moisture exchangers in artificial ventilation. An experimental study of the effect of gas leakage.

نویسندگان

  • S E Tilling
  • B Hayes
چکیده

The effect of gas leakage around an uncuffed tracheal tube on the performance of Heat and Moisture Exchangers (HME) was examined experimentally. Such leaks were shown to cause measurable decreases in the humidity output and efficiency of HME. The magnitude of the decrease in humidity output was shown to be dependent upon the proportion of expired volume which was leaked, and on the inspired volume. Other factors are discussed.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An Investigation on the Effects of Gas Pressure Drop in Heat Exchangers on Dynamics of a Free Piston Stirling Engine

This paper is devoted to study the effects of pressure drop in heat exchangers on the dynamics of a free piston Stirling engine. First, the dynamic equations governing the pistons as well as the gas pressure equations for hot and cold spaces of the engine are extracted. Then, by substituting the obtained pressure equations into the dynamic relationships the final nonlinear dynamic equations gov...

متن کامل

Effect of NOCOLOK® flux dry-off temperature on brazed joint strength of aluminum-based heat exchangers

During the brazing of aluminum-based heat exchangers, the flux dry-off temperature plays a crucial role to get sound joints with maximum strength. In the present study, the NOCOLOK® flux consists of two phases of K2AlF5.H2O and KAlF4 with a melting point around 580 °C was used for brazing AA3003 as base metal with a clad-coating of AA4343 as filler metal. The slurry was applied on the joints an...

متن کامل

Humidification on Ventilated Patients: Heated Humidifications or Heat and Moisture Exchangers?

The normal physiology of conditioning of inspired gases is altered when the patient requires an artificial airway access and an invasive mechanical ventilation (IMV). The endotracheal tube (ETT) removes the natural mechanisms of filtration, humidification and warming of inspired air. Despite the noninvasive ventilation (NIMV) in the upper airways, humidification of inspired gas may not be optim...

متن کامل

Humidification Performance of Heat and Moisture Exchangers for Pediatric Use

Background. While heat and moisture exchangers (HMEs) have been increasingly used for humidification during mechanical ventilation, the efficacy of pediatric HMEs has not yet been fully evaluated. Methods. We tested ten pediatric HMEs when mechanically ventilating a model lung at respiratory rates of 20 and 30 breaths/min and pressure control of 10, 15, and 20 cmH(2)O. The expiratory gas passed...

متن کامل

Ventilatory changes during the use of heat and moisture exchangers in patients submitted to mechanical ventilation with support pressure and adjustments in ventilation parameters to compensate for these possible changes: a self-controlled intervention study in humans

OBJECTIVE To evaluate the possible changes in tidal volume, minute volume and respiratory rate caused by the use of a heat and moisture exchanger in patients receiving pressure support mechanical ventilation and to quantify the variation in pressure support required to compensate for the effect caused by the heat and moisture exchanger. METHODS Patients under invasive mechanical ventilation i...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • British journal of anaesthesia

دوره 59 9  شماره 

صفحات  -

تاریخ انتشار 1987