Validation of TM-2655 oscillometric device for blood pressure measurement.
نویسندگان
چکیده
OBJECTIVE To perform validation for an arm-type oscillometric TM-2655 device (A&D Company Ltd, Tokyo, Japan) for blood pressure measurement according to the British Hypertension Society protocol. METHODS Eighty-five study participants (33 men and 52 women) were included in the study. Mean age was 52.9+/-15.0 years, systolic blood pressure range was 84-208 mmHg and diastolic blood pressure range was 48-120 mmHg. For each participant, three readings of TM-2655 were compared with sequential auscultatory measurements by two trained independent observers. The observers used a calibrated mercury sphygmomanometer and dual stethoscope. The results were graded according to the British Hypertension Society protocol 1993. RESULTS The average difference between mercury sphygmomanometer and TM-2655 readings for systolic blood pressure was -1.0+5.2 mmHg (mean+/-SD) and for diastolic blood pressure -0.9+/-4.7 mmHg. The proportions of values agreeing to within 5, 10 and 15 mmHg were 72.5, 93.7 and 99.6% for systolic blood pressure and 78.8, 96.9 and 100% for diastolic blood pressure between the observers and the device (A/A British Hypertension Society grade). CONCLUSIONS The TM-2655 device achieved British Hypertension Society grade A/A and therefore can be recommended for blood pressure measurement in an adult population.
منابع مشابه
Oscillometric blood pressure measurement: the methodology, some observations, and suggestions.
Most noninvasive blood pressure (NIBP) devices use the oscillometric method. Published studies of oscillometric methodology introduced varied algorithmic approaches for determination of systolic (SBP), diastolic (DBP), and mean arterial (MAP) blood pressures. While there is a general agreement about MAP determination, controversy exists about the determination of SBP and DBP Accuracy of oscillo...
متن کاملQuality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation
This study aims to validate blood pressure (BP) values measured by an oscillometric BP monitor and seek possible calibration methods if discrepancies exist. Noninvasive BP measurement outcomes were determined using an oscillometric BP monitor (Omron HBP-1300) versus a mercury sphygmomanometer (standard device). Two percent of subjects enrolled in an epidemiological investigation were systematic...
متن کاملComparison of methods of automatic blood pressure measurement in the same device
Comparison of tacho-oscillographic and oscillometric methods for human blood pressure measurement with cuff occlusion in the same device at the same time is described. For this purpose the measurement system and software for signal processing that realize these methods in the same device was developed. Experiment shows that oscillometric method with photoplethysmographic (PPG) sensors does not ...
متن کاملAssessment of the DIXTAL DX-2710 automated oscillometric device for blood pressure measurement with the validation protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI).
OBJECTIVE To assess the Dixtal DX2710 automated oscillometric device used for blood pressure measurement according to the protocols of the BHS and the AAMI. METHODS Three blood pressure measurements were taken in 94 patients (53 females 15 to 80 years). The measurements were taken randomly by 2 observers trained to measure blood pressure with a mercury column device connected with an automate...
متن کاملProspective clinical study to evaluate an oscillometric blood pressure monitor in pet rabbits
BACKGROUND Rabbits are particularly sensitive to develop hypotension during sedation or anaesthesia. Values of systolic or mean non-invasive arterial blood pressure below 80 or 60 mmHg respectively are common under anaesthesia despite an ongoing surgery. A reliable method of monitoring arterial blood pressure is extremely important, although invasive technique is not always possible due to the ...