BACKGROUND While navigation systems have been developed to increase implantation accuracy in total hip arthroplasty (THA), they are not yet sufficiently versatile or commonly used. Therefore, to elevate the appeal of such systems, we have developed HipCOMPASS, a simple and effective mechanical angle indicator for use in supine THA. QUESTIONS/PURPOSES How accurate is the mean cup orientation [in terms of errors in radiographic anteversion (RA) and inclination (RI)] in cases where HipCOMPASS is used for intraoperative support? Does HipCOMPASS increase this cup orientation accuracy compared to THA cases without it? Does HipCOMPASS increase mean operation time? METHODS We measured cup orientation in 97 THA cases with HipCOMPASS and in 80 cases without it. Then we compared the angles determined in preoperative planning with the angles revealed by postoperative computed tomography (CT) for both groups. The discrepancy between them was defined as an error. Errors greater than 10° were considered outliers. Additionally, mean operative time with and without the Hip COMPASS were compared. RESULTS With the use of HipCOMPASS, the mean absolute error values in radiographic anteversion and inclination were 2.9° ± 2.3° (range 0°-12.8°) and 2.9° ± 2.1° (0.1°-7.7°), respectively. In contrast, without the use of HipCOMPASS, radiographic anteversion and inclination error values were 8.8° ± 5.8° (0.1°-25.4°) and 6.1° ± 4.5° (0.2°-21.0°), respectively. Outlier occurrence rates were 1.0 % with HipCOMPASS and 48.8 % without it. Mean operative times with and without HipCOMPASS use were 109.2 ± 23.8 min (74-199 min) and 137.6 ± 40.6 min (71-298 min), respectively. CONCLUSIONS The study has found that HipCOMPASS dramatically increases implantation accuracy and it is also a simple and highly versatile tool that can be implemented quickly. Given its low cost in addition to its favourable accuracy, simple implementation, and short operative time, HipCOMPASS can be regarded as a very useful and effective THA support device. LEVEL OF EVIDENCE Retrospective comparative study, Level 3.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Visual intraoperative estimation of cup and stem position is not reliable in minimally invasive hip arthroplasty

Background and purpose - In hip arthroplasty, acetabular inclination and anteversion-and also femoral stem torsion-are generally assessed by eye intraoperatively. We assessed whether visual estimation of cup and stem position is reliable. Patients and methods - In the course of a subgroup analysis of a prospective clinical trial, 65 patients underwent cementless hip arthroplasty using a minimal...

متن کامل

Development and evaluation of an image-free computer-assisted impingement detection technique for total hip arthroplasty.

Periprosthetic or bony impingement in total hip arthroplasty (THA) has been correlated to dislocation, increased wear, reduced postoperative functionality with pain and/or decreased range of motion (ROM). We sought to study the accuracy and assess the reliability of measuring bony and periprosthetic impingement on a virtual bone model prior to the implantation of the acetabular cup with the hel...

متن کامل

Improving cup positioning using a mechanical navigation instrument.

BACKGROUND Although surgical navigation reduces the rate of malpositioned acetabular cups in total hip arthroplasty (THA), its use has not been widely adopted. As a result of our perceived need for simple and efficient methods of navigation, we developed a mechanical navigation device for acetabular cup orientation. QUESTIONS/PURPOSES We assessed accuracy of cup orientation (mean error of cup...

متن کامل

Accuracy of Image-based Computer Assisted Surgery in Acetabular Cup Placement for Supine Two-incision Total Hip Arthroplasty

Introduction: Supine patient positioning in conjunction with twoincision minimally invasive surgery (MIS-2 incision) total hip arthroplasty (THA) increases difficulty of accurate and repeatable cup orientation in accordance with pre-operative planning. Investigators have suggested advantages of utilizing computer assisted orthopaedic surgery (CAOS) to successfully eliminate extreme outlying ace...

متن کامل

Use of image-free navigation in determination of acetabular cup orientation: analysis of factors affecting precision.

We have been using an image-free total hip arthroplasty (THA) navigation system (OrthoPilot; Aesculap, Tuttlingen, Germany) to ensure accurate and reproducible acetabular cup orientation. In this study, the accuracy of this system in the assessment of acetabular cup orientation was evaluated by comparing the intraoperative inclination and anteversion angles presented by the navigation system an...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 5  شماره 

صفحات  -

تاریخ انتشار 2016