Effects of left ventricular unloading by Impella recover LP2.5 on coronary hemodynamics.
نویسندگان
چکیده
OBJECTIVES We studied the effects of LV unloading by the Impella on coronary hemodynamics by simultaneously measuring intracoronary pressure and flow and the derived parameters fractional flow reserve (FFR), coronary flow velocity reserve (CFVR), and coronary microvascular resistance (MR). BACKGROUND Patients with compromised left ventricular (LV) function undergoing high-risk percutaneous coronary intervention (PCI) may benefit from LV unloading. Limited information is available on the effects of LV unloading on coronary hemodynamics. METHODS Eleven patients (mean LV ejection fraction of 35 +/- 11%) underwent PCI during LV support by the LV unloading device (Impella Recover LP2.5). Intracoronary measurements were performed in a nonstenotic coronary artery after the PCI, before and after adenosine-induced hyperemia at four different support levels (0-2.5 L/min). RESULTS Aortic and coronary pressure increased with increasing support levels, whereas FFR remained unchanged. Baseline flow velocity remained unchanged, while hyperemic flow velocity and CFVR increased significantly with increasing support levels (61 +/- 24 to 72 +/- 27 cm/sec, P = 0.001 and 1.88 +/- 0.52 to 2.34 +/- 0.63, P < 0.001 respectively). The difference between baseline MR and hyperemic MR significantly increased with increasing support levels (1.28 +/- 1.32 to 1.89 +/- 1.43 mm Hg cm(-1) sec, P = 0.005). CONCLUSIONS Unloading of the LV by the Impella increased aortic and intracoronary pressure, hyperemic flow velocity and CFVR, and decreased MR. The Impella-induced increase in coronary flow, probably results from both an increased perfusion pressure and a decreased LV volume-related intramyocardial resistance.
منابع مشابه
Role of percutaneous coronary minimalist intervention in the management of acute ST-segment elevation myocardial infarction.
1. Smalling RW, Cassidy DB, Barrett R, et al. Improved regional myocardial blood flow, left ventricular unloading, and infarct salvage using an axial-flow, transvalvular left ventricular assist device. A comparison with intra-aortic balloon counterpulsation and reperfusion alone in a canine infarction model. Circulation 1992;85:1152–9. 2. Henriques JP, Remmelink M, Baan J Jr., et al. Safety and...
متن کاملA randomized clinical trial to evaluate the safety and efficacy of a percutaneous left ventricular assist device versus intra-aortic balloon pumping for treatment of cardiogenic shock caused by myocardial infarction.
OBJECTIVES The aim of this study was to test whether the left ventricular assist device (LVAD) Impella LP2.5 (Abiomed Europe GmbH, Aachen, Germany) provides superior hemodynamic support compared with the intra-aortic balloon pump (IABP). BACKGROUND Cardiogenic shock caused by left ventricular failure is associated with high mortality in patients with acute myocardial infarction (AMI). An LVAD...
متن کاملLeft Ventricular Mechanical Support with the Impella during Extracorporeal Membrane Oxygenation
Background: Venoarterial extracorporeal membrane oxygenation (ECMO) provides systemic arterial support without directly unloading the left heart, which causes an elevated left ventricular (LV) pressure. The aim of the present study was to investigate the adjunctive application of the Impella device for LV unloading in patients during ECMO. Methods: This retrospective cohort study included patie...
متن کاملIn vitro comparison of support capabilities of intra-aortic balloon pump and Impella 2.5 left percutaneous.
The Impella 2.5 left percutaneous (LP), a relatively new transvalvular assist device, challenges the position of the intra-aortic balloon pump (IABP), which has a long record in supporting patients after myocardial infarction and cardiac surgery. However, while more costly and more demanding in management, the advantages of the Impella 2.5 LP are yet to be established. The aim of this study was...
متن کامل[Initial experience with the Impella left ventricular assist device for postcardiotomy cardiogenic shock and unprotected left coronary artery angioplasty in patients with a low left ventricular ejection fraction].
Low-output syndrome is one of the leading causes of death following open-heart surgery or high-risk angioplasty. Ventricular assist devices have been used to treat patients who suffer from postoperative cardiogenic shock despite use of an intraaortic balloon pump and maximum inotropic support. The Impella pump (Impella Cardiosystems AG, Aachen, Germany) is a newly introduced left ventricular as...