The wear properties of CFR-PEEK-OPTIMA articulating against ceramic assessed on a multidirectional pin-on-plate machine.
نویسندگان
چکیده
In an attempt to prolong the lives of rubbing implantable devices, several 'new' materials have been examined to determine their suitability as joint couplings. Tests were performed on a multidirectional pin-on-plate machine to determine the wear of both pitch and PAN (polyacrylonitrile)-based carbon fibre reinforced-polyetheretherketone (CFR-PEEK-OPTIMA) pins articulating against both BioLox Delta and BioLox Forte plates (ceramic materials). Both reciprocation and rotational motion were applied to the samples. The tests were conducted using 24.5 per cent bovine serum as the lubricant (protein concentration 15 g/l). Although all four material combinations gave similar low wear with no statistically significant difference (p > 0.25), the lowest average total wear of these pin-on-plate tests was provided by CFR-PEEK-OPTIMA pitch pins versus BioLox Forte plates. This was much lower than the wear produced by conventional joint materials (metal-on-polyethylene) and metal-on-metal combinations when tested on the pin-on-plate machine. This therefore indicates optimism that these PEEK-OPTIMA-based material combinations may perform well in joint applications.
منابع مشابه
Influence of contact pressure, cross-shear and counterface material on the wear of PEEK and CFR-PEEK for orthopaedic applications
Total joint replacement is a successful surgical intervention for the treatment of the degeneration of many joints, particularly the hip and knee. As the demand for joint replacement grows, and the life expectancy of the population increases, the performance requirements of these implants also changes. New materials, to improve longevity and enhance performance have been explored including PEEK...
متن کاملPEEK and CFR-PEEK as alternative bearing materials to UHMWPE in a fixed bearing total knee replacement: An experimental wear study
New bearing materials for total joint replacement have been explored as the need to improve longevity and enhance performance is driven by the changing demands of the patient demographic. Carbon-reinforced PEEK has demonstrated good wear characteristics in experimental wear simulation in both simple geometry pin-on-plate studies and in total hip joint replacement. Carbon reinforced PEEK CFR-PEE...
متن کاملPEEK-OPTIMA™ as an alternative to cobalt chrome in the femoral component of total knee replacement: A preliminary study
PEEK-OPTIMA™ (Invibio Ltd, UK) has been considered as an alternative joint arthroplasty bearing material due to its favourable mechanical properties and the biocompatibility of its wear debris. In this study, the potential to use injection moulded PEEK-OPTIMA™ as an alternative to cobalt chrome in the femoral component of a total knee replacement was investigated in terms of its wear performanc...
متن کاملWear of ceramic-on-carbon fibre reinforced PEEK hip replacements
Total hip replacement has been a successful surgical intervention for many years. However, concerns regarding UHMWPE wear particle induced osteolysis [1], have led to alternative bearing materials to be sought. Carbon-fibre reinforced poly-ether-ether-ketone (CFR-PEEK) is a fibre-reinforced polymer which has shown reduced wear in hip and knee configurations compared with conventional polyethyle...
متن کاملWear and Wear Debris Generated from Ceramic-on-Carbon Fibre Reinforced PEEK Hip Replacements
INTRODUCTION Concerns regarding UHMWPE wear particle induced osteolysis [1] have led to investigations of alternative bearing materials. Carbon-fibre reinforced poly-ether-ether-ketone (CFR-PEEK) has been shown in animal studies to have a similar biological activity to UHMWPE [2], whilst hip and knee configurations incorporating this polymer display reduced wear rates when compared with convent...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
دوره 221 3 شماره
صفحات -