Electron beam central axis depth dose measurements.
نویسندگان
چکیده
Central axis depth dose measurements were made by the Radiological Physics Center on over 70 electron-producing machines used in radiation therapy. These data were consistent for each machine model and nominal energy. However, the data show that depth dose relations can vary significantly among different machine models for electron beams having the same nominal energy. Analysis shows that both the method used to achieve beam flatness and the mean incident electron energy determine the central axis depth dose curve past the depth of maximum dose. A linear relation of depth dose versus mean incident electron energy is used to predict depth dose to within 2 mm for most electron beams used clinically at depths greater than d95.
منابع مشابه
Experimental assessment of out‐of‐field dose components in high energy electron beams used in external beam radiotherapy
The purpose of this work was to experimentally investigate the out-of-field dose in a water phantom, with several high energy electron beams used in external beam radiotherapy (RT). The study was carried out for 6, 9, 12, and 18 MeV electron beams, on three different linear accelerators, each equipped with a specific applicator. Measurements were performed in a water phantom, at different depth...
متن کاملCommissioning the First Mobile Dedicated Accelerator for Intraoperative Electron Radiotherapy in Iran
Introduction: Intraoperative radiotherapy is a radiotherapy technique in which a high single fraction of radiation dose is delivered to the patient after surgery and Concurrent with anesthesia time. The most frequent method for IORT implementation is Intraoperative electron radiotherapy (IOERT), in which, some dedicated and high dose per pulse electron accelerators are employe...
متن کاملElectron Beam Dose Distribution in the Presence of Non-Uniform Magnetic Field
Introduction Magnetic fields are capable of altering the trajectory of electron beams andcan be used in radiation therapy.Theaim of this study was to produce regions with dose enhancement and reduction in the medium. Materials and Methods The NdFeB permanent magnets were arranged on the electron applicator in several configurations. Then, after the passage of the electron beams (9 and 15 MeV Va...
متن کاملBenchmarking of Siemens Linac in Electron Modes: 8-14 MeV Electron Beams
Introduction: Radiation therapy using electron beams is a promising method due to its physical dose distribution. Monte Carlo (MC) code is the best and most accurate technique for forespeaking the distribution of dose in radiation treatment of patients.Materials and Methods: We report an MC simulation of a linac head and depth dose on central axis, along with profile calculations. The purpose o...
متن کاملEffect of Tissue Composition on Dose Distribution in Electron Beam Radiotherapy
Objective: The aim of this study is to evaluate the effect of tissue composition on dose distribution in electron beam radiotherapy.Methods: A Siemens Primus linear accelerator and a phantom were simulated using MCNPX Monte Carlo code. In a homogeneous cylindrical phantom, six types of soft tissue and three types of tissue-equivalent materials were investigated. The tissues included muscle (ske...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Medical physics
دوره 12 3 شماره
صفحات -
تاریخ انتشار 1985