Validation of Missouri aneroid sphygmomanometer to measure blood pressure in patients with cancer.

نویسندگان

  • Karine Azevedo São Leão Ferreira
  • Ana Claúdia dos Santos
  • Thais Cardoso Arthur
  • Daniela Aparecida A Dos Santos
  • Daniela Pereira
  • Elizângela Oliveira Freitas
  • Fernanda Medeiros Fukuda
  • Wânia Regina Mollo Baia
چکیده

BACKGROUND Auscultatory mercury sphygmomanometers to measure blood pressure (BP) have been banned from health services because of risk of pollution and environmental accidents with mercury. Aneroid appliances could be an alternative. OBJECTIVE To validate the Missouri aneroid device for blood pressure measurement in cancer patients according to the protocol of the European Society of Hypertension (ESH). METHODS 33 patients hospitalized or under outpatient care at the Cancer Institute of the State of São Paulo, FMUSP, were evaluated. Three trained and blinded observers performed nine sequential blood pressure measurements interspersed with the mercury sphygmomanometers. The differences between the values of systolic blood pressure (SBP) and diastolic blood pressure (DBP) of the test device with the mercury sphygmomanometer were classified according to the ESH protocol. RESULTS The Missouri equipment underwent all three phases required by the ESH Protocol for SBP and DBP, and it was approved in all of the phases. The average difference between the test device and the mercury sphygmomanometer was 0.62 (SD = 4.53) and 0.06 (SD = 6.57) mmHg for SBP and DBP, respectively. No association was found between the differences in BP measurements with sex, age, body mass index and arm circumference and length. CONCLUSION The results revealed that the aneroid Missouri device meets ESH accuracy recommendations for the measurement of SBP and DBP, and it can be used to replace the mercury sphygmomanometer.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluating the accuracy of an aneroid sphygmomanometer in a clinical trial setting.

BACKGROUND The mercury sphygmomanometer, the "gold" standard for blood pressure measurements, has been gradually phased out in many institutions because of environmental concerns. Our on-going clinical trial compared the use of mercury vs. aneroid sphygmomanometers, before implementing a study-wide transition to the aneroid sphygmomanometer. METHODS The Diabetes Prevention Program Outcomes St...

متن کامل

Validation of Heine Gamma G7 (G5) and XXL-LF aneroid devices for blood pressure measurement.

OBJECTIVE To determine the accuracy of the Gamma G7 (and G5 model) and XXL-LF aneroid sphygmomanometers developed by the Heine Company. DESIGN Device evaluations were performed using the new protocol of the European Society of Hypertension. Monitor performance was assessed in relation to participants' sex, age, arm circumference, and systolic and diastolic blood pressures. METHODS The two s...

متن کامل

Measuring blood pressure at the wrist: more comfortable for patients and more convenient for doctors?

OBJECTIVES To compare the agreement between conventional measurement of blood pressure and measurements obtained using two automated devices; and to compare how comfortable each of the three methods of measurement were for patients. METHODS Blood pressure measurements and patient comfort scores were recorded using three different devices in 125-surgery and 40-community patients. The devices u...

متن کامل

بررسی کیفیت سنجش فشار خون توسط پرستاران مجتمع بیمارستانی امام خمینی تهران سال 1389

Background & aims: Blood pressure is an important sign of cardiovascular, renal and cerebral system function and is the base of making important medical decision. However it is nurses' routine duty, accurate measurement is received little attention. Therefore this study was done to evaluate the quality of blood pressure measuring by nurses. Methods: This is a cross-sectional descriptive st...

متن کامل

مقایسه اندازه گیری فشارخون توسط گوشی معمولی با گوشی تلفیقی

Background and Purpose: Blood pressure (BP) is routinely evaluated intermittently during the clinical status for care and treatment. The most common method to measure BP is by using the sphygmomanometer with a stethoscope. This method may be difficult for medical students that will use BP for medical evaluation. The purpose of this study was to compare the accuracy of double stethoscope devic...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Arquivos brasileiros de cardiologia

دوره 95 2  شماره 

صفحات  -

تاریخ انتشار 2010