Analytical Evaluation of the Clinical Chemistry Analyzer AU 480 Beckman Coulter
نویسنده
چکیده
Background: Method verification is a one-time process to determine performance characteristics before a test system is utilized for patient testing. Objective: To evaluate the analytical performance of five analytes-glucose (glu), cholesterol (chol), creatinine (crea), aspartate aminotransferase, (AST) and alanine aminotransferase (ALT) on AU 480 Beckman Coulter clinical chemistry analyzer. Methods and Materials: Analytical evaluation of analyzer included the determination of within-run, within-laboratory imprecision, and Trueness. Beckman Coulter control level-one and control level-two, near medical decision points were used. It was done according to (CLSI) clinical laboratory standard institute (EP15-A2), which suggests two levels, run 3 times per run for 5 days (15replicates in all). Result: The results showed low within-run imprecision (level 1-ALT=1.5%, AST=1.3%, chol=0.5%, crea=0.2%, glu=0.7%, level 2ALT=1.0%, AST=1.3%, chol=0.6%, crea=0.9%, glu=0.4%) and within-laboratory imprecision (level 1-ALT=2.6%, AST=0.9%, chol=0.7%, crea=0.2%, glu=0.8%, level 2ALT=1.9%, AST=1.0%, chol=0.6%, crea=0.8%, glu=0.5%)and the assigned values were found to be within trueness verification intervals (level 1 at 95% CI: ALT=46.5U/L[22.7-66.6], AST=50.4U/L[29.1-76.5], chol=156mg/dl[107.1197.1], crea=1.19mg/dl [0.96-2.03], glu=103mg/dl [69.8-136.3]) and (level 2 at 95% CI: ALT=133U/L [60.5-187.3], AST=148U/L [81.3-220.3], chol=285mg/dl [186.2-349.8], crea=5.35mg/dl [3.46-8.3], glu=239mg/dl [163-320.5]). Conclusion: AU 480 Beckman Coulter clinical chemistry analyzer is suitable for analytical measurement of analytes-AST, ALT, cholesterol, creatinine, and glucose. If laboratories use CLSI and other guidelines, which are published on issues of biological variations, in addition to the manufacturer’s claims, it may help laboratories not to waste unnecessary time and money for repeating of experiment.
منابع مشابه
Evaluation of NGAL Test™, a fully-automated neutrophil gelatinase-associated lipocalin (NGAL) immunoassay on Beckman Coulter AU 5822.
BACKGROUND The neutrophil gelatinase associated lipocalin (NGAL) has been identified as the most promising biomarker of acute kidney injury (AKI). This study was aimed to evaluate a NGAL immunoassay on Beckman Coulter AU 5822. METHODS NGAL Test™ (BioPorto Diagnostics A/S) is a particle-enhanced turbidimetric immunoassay. The within-and between-run imprecision were assessed on three urine samp...
متن کاملMulti-center evaluation of analytical performance of the Beckman Coulter AU5822 chemistry analyzer.
OBJECTIVES Our three academic institutions, Indiana University, Northwestern Memorial Hospital, and Wake Forest, were among the first in the United States to implement the Beckman Coulter AU5822 series chemistry analyzers. We undertook this post-hoc multi-center study by merging our data to determine performance characteristics and the impact of methodology changes on analyte measurement. DES...
متن کاملAn Evaluation of the AcT 5diff Cap Pierce Hematology Analyzer
Staffing constraints and cross-training of laboratory staff have made it imperative that we choose laboratory instruments that not only are user friendly and easy to use but also improve the efficiency of our current systems. Therefore, evaluations of any instrument should also incorporate a component that examines whether workflow can be improved by the use of the instrument being evaluated. I...
متن کاملEvaluation of the Beckman Coulter UniCel DxH 800, Beckman Coulter LH 780, and Abbott Diagnostics Cell-Dyn Sapphire hematology analyzers on adult specimens in a tertiary care hospital.
We evaluated the new Beckman Coulter DxH 800 hematology analyzer (Beckman Coulter, Miami, FL) vs the Abbott Diagnostics Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA) and Beckman Coulter LH 780 hematology analyzers using 430 adult specimens. The DxH 800 provided a CBC and differential that correlated well with those of the Sapphire and LH 780, with most parameters showing correlation c...
متن کاملInfluence of blood sampling site on intact parathyroid hormone concentrations in hemodialysis patients.
Employment or Leadership: None declared. Consultant or Advisory Role: P.A. Kavsak, Ortho Clinical Diagnostics; A.R. MacRae, Beckman Coulter; A.S. Jaffe, Beckman Coulter, Siemens, Ortho Diagnostics, Singulex, Nanosphere, Inverness, Critical Diagnostics, GSK, Merck, and Hoffman-LaRoche. Stock Ownership: None declared. Honoraria: P.A. Kavsak, Beckman Coulter. Research Funding: Canadian Institutes ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2016