We tested Portapres, an innovative portable, battery-operated device for the continuous, noninvasive, 24-hour ambulatory measurement of blood pressure in the finger. Portapres is based on Finapres, a stationary device for the measurement of finger arterial pressure. Systems were added to record signals on tape, to alternate measurements between fingers automatically each 30 minutes, and to correct for the hydrostatic height of the hand. We compared the pressure as measured by Portapres with contralateral intrabrachial pressure measured with an Oxford device. Results were obtained in eight volunteers and 16 hypertensive patients. Time lost due to artifact was about 10% for each device. In two patients a full 24-hour Oxford profile was not obtained. In the remaining 22 subjects finger systolic, diastolic, and mean pressures differed +1 (SD 9), -8 (6), and -10 (6) mm Hg, respectively, from intrabrachial pressure. These diastolic and mean pressure underestimations are similar to what was found earlier for Finapres, are typical for the technique, and are systematic. Avoiding brisk hand movements resulted in fewer waveform artifacts. The hand had to be kept covered to continue recording at low outside temperatures. Sleep was not disturbed by Portapres, and arterial pressure showed a marked fall during siesta and nighttime. There were no major limitations in behavior, and no discomfort that originated from continuous monitoring was reported. Measurements continued normally during physical exercise. Portapres provides for the first time continuous 24-hour, noninvasive ambulatory blood pressure waveform monitoring and offers real and obvious advantages over current noninvasive and invasive devices.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure.

This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial dias...

متن کامل

Estimation of blood pressure variability from 24-hour ambulatory finger blood pressure.

Portapres is a noninvasive, beat-to-beat finger blood pressure (BP) monitor that has been shown to accurately estimate 24-hour intra-arterial BP at normal and high BPs. However, no information is available on the ability of this device to accurately track ambulatory BP variability. In 20 ambulatory normotensive and hypertensive subjects, we measured 24-hour BP by Portapres and through a brachia...

متن کامل

Feasibility of using finger arterial pressure in neonates.

The feasibility of using a Finapres device to reproduce the beat to beat signal of arterial blood pressure in eight neonates was assessed and compared with intra-arterial measurement of arterial blood pressure in the umbilical artery, using a catheter. The two methods gave similar results. Continuous recording of arterial blood pressure in neonates using Finapres is feasible and reliable.

متن کامل

Continuous recording of direct arterial pressure in unrestricted patients-its role in the diagnosis and management of high blood pressure.

We have compared casual indirect measurements of arterial pressure obtained 1) by the general practitioner (GP) and 2) in the outpatient clinic (OPC) with 24 hour continuous recording of direct arterial pressure in two selected group of unrestricted patients. 1) Eight asymptomatic, untreated patients with suspected hypertension. 2) Eight asymptomatic, treated patients whose indirect pressure re...

متن کامل

Prediction of coronary and cerebrovascular morbidity and mortality by direct continuous ambulatory blood pressure monitoring in essential hypertension.

BACKGROUND The goal of this study was to assess the prognostic value of ambulatory versus clinic blood pressure measurement and to relate cardiovascular risk to ambulatory systolic and diastolic blood pressure levels. METHODS AND RESULTS The study population consisted of 688 patients 51+/-11 years of age who had undergone pretreatment 24-hour intra-arterial ambulatory blood pressure monitorin...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Hypertension

دوره 21 1  شماره 

صفحات  -

تاریخ انتشار 1993