Direct EIT Jacobian calculations for conductivity change and electrode movement.
نویسندگان
چکیده
Electrical impedance tomography (EIT) is very sensitive to deformations of the medium boundary shape. For lung imaging, breathing and changes in posture move the electrodes and change the chest shape, resulting in image artefacts. Several approaches have been proposed to improve the reconstructed images; most methods reconstruct both the boundary deformation and conductivity change from the measured data. These techniques require the calculation of the 'movement Jacobian', reflecting measurement changes due to the boundary deformation. Previous papers have calculated this Jacobian using perturbation techniques, which are slow (requiring multiple solutions of the forward problem) and become inaccurate with increasing finite element model size. This effect has limited reconstruction algorithms for deformable media to mostly 2D. To address this problem, we propose a direct method to calculate the Jacobian, based on a formulation of the derivatives of the finite element system matrix with respect to geometry changes. An illustrative example of these calculations is given, as well as a comparison between the proposed method and a perturbation method. Results show this method is approximately 300 times faster; and for larger model sizes, the perturbation method begins to diverge from those from the direct method proposed.
منابع مشابه
Direct calculation of the electrode movement Jacobian for 3D EIT
Electrical Impedance Tomography (EIT) of media with deformable boundaries is very sensitive to electrode movement. This is especially important for EIT images of the thorax, which become distorted with breathing and posture change. Previously, we proposed a reconstruction method for imaging conductivity change and electrode movement based on an indirect perturbation Jacobian calculation, involv...
متن کاملImaging of conductivity changes and electrode movement in EIT.
Electrical impedance tomography (EIT) attempts to reconstruct the internal impedance distribution in a medium from electrical measurements at electrodes on the medium surface. One key difficulty with EIT measurements is due to the position uncertainty of the electrodes, especially for medical applications, in which the body surface moves during breathing and posture change. In this paper, we de...
متن کاملReconstruction of conductivity changes and electrode movements based on EIT temporal sequences.
Electrical impedance tomography (EIT) reconstructs a conductivity change image within a body from electrical measurements on the body surface; while it has relatively low spatial resolution, it has a high temporal resolution. One key difficulty with EIT measurements is due to the movement and position uncertainty of the electrodes, especially due to breathing and posture change. In this paper, ...
متن کاملTracking boundary movement and exterior shape modelling in lung EIT imaging.
Electrical impedance tomography (EIT) has shown significant promise for lung imaging. One key challenge for EIT in this application is the movement of electrodes during breathing, which introduces artefacts in reconstructed images. Various approaches have been proposed to compensate for electrode movement, but no comparison of these approaches is available. This paper analyses boundary model mi...
متن کاملCorrection of electrode modelling errors in multi-frequency EIT imaging.
The differentiation of haemorrhagic from ischaemic stroke using electrical impedance tomography (EIT) requires measurements at multiple frequencies, since the general lack of healthy measurements on the same patient excludes time-difference imaging methods. It has previously been shown that the inaccurate modelling of electrodes constitutes one of the largest sources of image artefacts in non-l...