Inhalational drug delivery from seven different spacer devices.
نویسندگان
چکیده
BACKGROUND A study was performed to determine in vitro the difference in drug output of seven currently available spacer devices when used with different inhaled medications. METHODS A glass multistage liquid impinger (MSLI) was used to determine the amount of disodium cromoglycate (DSCG, 5 mg), salbutamol (100 micrograms), or budesonide (200 micrograms) obtained in various particle size ranges from metered dose inhalers (MDIs) actuated directly into the MSLI or via one of seven different spacer devices; the Fisonair, Nebuhaler, Volumatic, Inspirease, Aerochamber, Aerosol Cloud Enhancer, and Dynahaler. RESULTS In particles smaller than 5 microns in diameter the dose of DSCG recovered from the Fisonair and Nebuhaler was 118% and 124%, respectively, of that recovered using the MDI alone. The dose recovered from the smaller volume spacers was 90% (Inspirease), 36% (Aerochamber), 33% (Aerosol Cloud Enhancer), and 21% (Dynahaler) of that from the MDI alone. The Volumatic increased the amount of salbutamol in particles smaller than 5 microns to 117% of that from the MDI, and the Inspirease and Aerochamber spacers decreased it by nearly 50%. The amount of budesonide in small particles recovered after use of the Nebuhaler, Inspirease, and the Aerochamber was 92%, 101%, and 78%, respectively, of that from the MDI alone. CONCLUSIONS Under the test conditions used, large volume spacers such as the Fisonair, Nebuhaler, and Volumatic delivered significantly more DSCG and salbutamol than the smaller spacers tested. The differences between spacers were less for budesonide than the other medications studied. This study shows that there are significant differences in the amount of drug available for inhalation when different spacers are used as inhalational aids with different drugs. Spacer devices need to be fully evaluated for each drug prescribed for them.
منابع مشابه
In vitro comparison of the amount of salbutamol available for inhalation from different formulations used with different spacer devices.
Metered-dose inhalers (MDIs) are currently being reformulated to contain hydrofluoroalkanes (HFAs), which do not damage the Earth's ozone layer. As different formulations of inhaled drugs may behave differently when used with spacer devices, we wished to determine the amount of salbutamol available for inhalation from a conventional metered-dose inhaler (Ventolin) and a new HFA-containing formu...
متن کاملInfluence of spacer device on drug delivery to young children with asthma.
The budesonide dose delivered to the patient from three different spacer devices (Nebuhaler = 750 ml, Aerochamber = 140 ml, and Babyspacer = 260 ml) was assessed by measuring the budesonide dose deposited on a filter inserted between the spacer outlet and the mouth of the patient. Twenty children aged 10-25 months were given a single dose of 200 micrograms budesonide from each spacer device in ...
متن کاملThe Effect of Spacer Morphology on the Aerosolization Performance of Metered-Dose Inhalers.
PURPOSE Respiratory drug delivery has been attracted great interest for the past decades, because of the high incidence of pulmonary diseases. However, despite its invaluable benefits, there are some major drawbacks in respiratory drug delivery, mainly due to the relatively high drug deposition in undesirable regions. One way to improve the efficiency of respiratory drug delivery through metere...
متن کاملMetered dose inhalers and spacer devices.
Easy to carry, highly effective and extremely safe, allowing accurate, consistent dose delivery, metered dose inhalers are the inhalation devices doctors and patients choose most readily the world over. Clinical response, however, may be affected by the inhalation technique used and the precise formulation in the canisters. The purpose of this review is to consider metered dose inhalers, add on...
متن کاملThe effects of static charge in spacer devices on glucocorticosteroid aerosol deposition in asthmatic patients.
Electrostatic charge in plastic spacer devices has been shown in vitro to reduce delivery of asthma medications intended for inhalation, but the effect of static charge on in vivo drug deposition is unknown. A six-way randomized crossover study was conducted in 10 mild asthmatic patients. Two plastic spacers (Nebuhaler and Volumatic) and one metal spacer (Nebuchamber) were tested. The spacers w...