Determination of molybdenum in steel samples by ICP-AES after separation and preconcentration using nanometre-sized titanium dioxide
نویسندگان
چکیده
A sensitive and selective method has been developed for the determination of Mo in steel samples using a nanometre-sized titanium dioxide micro-column (20 mm 6 3.0 mm id) and inductively coupled plasma atomic emission spectrometry (ICP-AES). Under the optimized conditions (pH 1.0, flow rate 1.5 mL min), Mo was retained on the column and separated from the iron matrix, then eluted with 0.5 mol L NaOH and determined by ICP-AES. The adsorption capacity of nanometre TiO2 for Mo was found to be 2.01 mg g . The detection limit for Mo was 0.14 mg L, with an enrichment factor of 100, and the relative standard deviation (RSD) was 1.8% (n ~ 11, C ~ 50 mg L). The proposed method has been applied to the determination of Mo in certified reference steel samples with satisfactory results.
منابع مشابه
Selective Separation and Preconcentration of trace Amounts of Gallium in Water and Rice Samples using Cloud Point Extraction and Determination by Inductively Coupled Plasma-Atomic Emission Spectrometry
In the present study a cloud-point extraction process using non-ionic surfactant Triton X-114 for selective extraction of gallium from aqueous solutions was developed. The method is based on the complex formation of Ga (III) with N, N′ -bis (salycilidene)-1, 2-phenylenediamine (salophen) as a chelating agent in buffer media of pH 5. After phase separation and dilution of the surfactant-rich pha...
متن کاملAdsorption behavior of noble metal ions (Au, Ag, Pd) on nanometer-size titanium dioxide with ICP-AES.
The adsorption behavior of gold (Au), silver (Ag) and palladium (Pd) on nanometer-size titanium dioxide (NSTD) at low concentrations was studied using inductively coupled plasma atomic emission spectrometry (ICP-AES). A preconcentration procedure of the analytes was carried out using NSTD as a solid-phase extractant before their determination by ICP-AES. The optimum conditions for adsorption we...
متن کاملDetermination of cobalt in urine by FI-ICP-AES with online preconcentration.
A method for the preconcentration and determination of cobalt in human urine samples was developed. The online preconcentration and determination were attained using inductively coupled plasma atomic emission spectromety (ICP-AES) coupled to a flow injection (FI) method. Cobalt was retained on an Amberlite XAD-7 resin as cobalt-2-(5-bromo-2-pyridylazo)-5-diethylaminophenol complex at pH 9.5. Co...
متن کاملDetermination of cadmium in water by ICP-AES with on-line adsorption preconcentration using DPTH-gel and TS-gel microcolumns.
Two flow injection inductively coupled plasma atomic emission spectrometric methods for the preconcentration and determination of trace amounts of cadmium in sea-water and waste-water samples are described based on the adsorption of the metal ion on a micro-column placed in the injection valve of the FI manifold and packed with silica gel funtionalised with 1,5-bis(di-2-pyridyl) methylene thioc...
متن کاملApplication of high-surface-area ZrO2 in preconcentration and determination of 18 elements by on-line flow injection with inductively coupled plasma atomic emission spectrometry.
A flow-injection analysis (FIA) system incorporating a micro-column of ZrO2 has been used for the development of an on-line multi-element method for the simultaneous preconcentration and determination of Al, Bi, Cd, Co, Cr, Cu, Fe, Ga, In, Mn, Mo, Ni, Pb, Tl, V, Sb, Sn, and Zn by inductively coupled plasma atomic emission spectrometry (ICP-AES). The conditions for quantitative and reproducible ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2004