Heterophilic antibodies may be a cause of falsely low total IGF1 levels.

نویسندگان

  • M P Brugts
  • J G L M Luermans
  • E G W M Lentjes
  • N J van Trooyen-van Vrouwerff
  • F A L van der Horst
  • P H Th J Slee
  • S W J Lamberts
  • J A M L Janssen
چکیده

BACKGROUND A low serum total IGF1 is considered as a diagnostic indicator of GH deficiency (GHD) in the presence of hypopituitarism. Introduction of IRMA and chemiluminescent immunometric assay (CLIA) IGF1 immunoassays has introduced endogenous antibodies as a new source of interference. In general, this goes unnoticed and might lead to unnecessary diagnostic and therapeutic interventions. CASE A 56-year-old man was referred with a decline in physical performance, unexplained osteopenia, and weight loss of 3 kg over the past 8 months. Although clinical signs and symptoms were unremarkable, laboratory results pointed to secondary hypothyroidism and secondary hypogonadism. In addition, the serum total IGF1 level (CLIA; Siemens Medical Solutions Diagnostics) was in the low normal range. Two GH stimulation tests were performed, but these tests did not support the diagnosis GHD. Moreover, IGF1 bioactivity measured by the kinase receptor activation assay was normal. Interference of heterophilic antibodies was considered. After pretreatment with specific heterophilic blocking tubes that contain blocking reagents to eliminate heterophilic antibodies, serum-free thyroxine, testosterone, and IGF1 levels turned out to be normal. CONCLUSION To the best of our knowledge, we here describe the first case in the literature of a patient with low serum total IGF1 levels due to interference from heterophilic antibodies in the used IGF1 immunoassay. When confronted with low-IGF1 levels that do not fit the clinical picture, interference of heterophilic antibodies should be considered in the differential diagnosis.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Heterophilic antibodies: a problem for all immunoassays.

We verified that antibody-binding substances in serum that interfere in two-site immunoassays involving murine antibodies are heterophilic antibodies. Incubation of serum containing heterophilic antibodies and a murine monoclonal antibody to human choriogonadotropin (hCG) leads to formation of a series of soluble immune complexes. We investigated the recognition of hCG by reagent antibody in th...

متن کامل

Use of an in vivo biotinylated single-chain antibody as capture reagent in an immunometric assay to decrease the incidence of interference from heterophilic antibodies.

BACKGROUND Heterophilic antibodies are a common source of interference in immunometric assays. We tested the hypothesis that the incidence of such interference could be decreased by use of a recombinant in vivo-biotinylated single-chain antibody (scFv) as the capture reagent. METHODS We established three assays for carcinoembryonic antigen (CEA) with the capture antibody either chemically bio...

متن کامل

Heterophilic antibodies in human sera causing falsely increased results in the CA 125 immunofluorometric assay.

An in-house OC 125 monoclonal antibody-based "sandwich" immunofluorometric assay (IFMA) described previously (Clin Chem 1987; 33:2191-4) gave higher results for CA 125 in 37 of 123 serum samples than did a commercially available immunoradiometric assay (IRMA). Discordant results between the two assays became concordant when measurements of the samples were repeated with normal mouse serum (100 ...

متن کامل

The war on heterophilic antibody interference.

Human heterophilic antibodies may bind the animal antibodies used in an immunoassay and thus produce erroneous results. Such interference has been increasingly recognized as a diagnostic problem (1, 2). In this issue of Clinical Chemistry, Adel Ismail (3 ) discusses the two main issues of heterophilic antibody interference in immunometric assays: how to detect it and how to avoid it. This is an...

متن کامل

Misleading FT4 measurement: Assay-dependent antibody interference

Introduction: Commonly used free thyroxine (FT4) immunoassays can be falsely elevated due to interference causing misinterpreted thyroid function. We present two cases with high FT4 concentrations due to antibody interference. This study’s aim was to investigate the source of the FT4 immunoassay interference and possibility of its removal by two different techniques in order to correct the disc...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • European journal of endocrinology

دوره 161 4  شماره 

صفحات  -

تاریخ انتشار 2009