Comparison of a new optical biometer and a standard biometer in cataract patients

نویسنده

  • Pipat Kongsap
چکیده

BACKGROUND Cataract surgery is the most common surgical procedure in ophthalmology. Biometry data and accurate intraocular lens (IOL) calculations are very important in achieving the desired refractive outcomes. The aim of this study was to compare measurements using a new optical low coherence reflectometry (OLCR) biometer (OA-2000) and the gold standard partial coherence interferometry (PCI) optical biometer (IOLMaster 500). METHODS Ocular biometry of cataract patients were measured by the OA-2000 and IOLMaster 500 to compare keratometry (K), axial length (AL), anterior chamber depth (ACD), white-to-white (WTW) diameter, and IOL power using the SRK/T formula. RESULTS One hundred and two eyes of 68 cataract patients were evaluated with the two optical biometers. The mean values of the AL, K, ACD, and WTW differed very little (OCLR biometer, 23.12 mm, 44.50 diopters (D), 3.06, and 11.64 mm, respectively; PCI biometer, 23.18 mm, 44.6 D, 3.15, and 11.86 mm, respectively), but the differences were significant (all, p ≤ 0.05). The AL, K, and ACD showed excellent correlations (r = 0.999, 0.980, and 0.824, respectively; all p < 0.001); however, there was a weak correlation of the WTW diameter between the two devices (r = 0.256). The IOL powers using the SRK-T formula derived from both instruments were very similar, with an excellent correlation (r = 0.989). The mean difference between the two instruments was 0.32 D. CONCLUSIONS The OLCR biometer showed very a strong agreement with the standard PCI optical biometer for almost all ocular biometry measurements, except for the WTW diameter. TRIAL REGISTRATION TCTR20160614003; date 06/09/2016; 'retrospectively registered'.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Biometry and intraocular lens power calculation results with a new optical biometry device: comparison with the gold standard.

PURPOSE To evaluate the agreement in axial length (AL), keratometry (K), anterior chamber depth (ACD) measurements; intraocular lens (IOL) power calculations; and predictability using a new partial coherence interferometry (PCI) optical biometer (AL-Scan) and a reference (gold standard) PCI optical biometer (IOLMaster 500). SETTING Service d'Ophtalmologie, Hopital Bicêtre, APHP Université, Pa...

متن کامل

Comparison of corneal astigmatism measurements of 2 optical biometer models for toric intraocular lens selection.

PURPOSE To compare corneal astigmatism measurements obtained using a new optical biometer (AL-Scan) and a standard biometer based on partial coherence interferometry (IOLMaster 500) for toric intraocular lens (IOL) selection. SETTING Siriraj Hospital, Mahidol University, Bangkok, Thailand. DESIGN Prospective comparative study. METHODS Keratometry (K) readings were taken with the 2 biomete...

متن کامل

Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometer.

PURPOSE To compare the repeatability and reproducibility of ocular biometry and intraocular lens (IOL) power obtained with a new optical biometer (AL-Scan) and a standard optical biometer (IOLMaster 500). SETTING Siriraj Hospital, Mahidol University, Bangkok, Thailand. DESIGN Prospective comparative study. METHODS Two independent operators measured eyes with cataract using both biometers....

متن کامل

Accuracy of optical biometry combined with Placido disc corneal topography for intraocular lens power calculation

PURPOSE To investigate the accuracy of a new optical biometer for intraocular lens (IOL) power calculation in eyes undergoing cataract surgery. METHODS Consecutive eyes of patients undergoing cataract surgery with the same IOL model were enrolled in a prospective cohort study. Axial length (AL) and corneal power were measured with an optical biometer based on optical low-coherence interferome...

متن کامل

Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology

PURPOSE The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 3  شماره 

صفحات  -

تاریخ انتشار 2016