The Effect of Compressor-administered Defibrillation on Peri-shock Pauses in a Simulated Cardiac Arrest Scenario

نویسندگان

  • Joshua Glick
  • Erik Lehman
  • Thomas Terndrup
چکیده

INTRODUCTION Coordination of the tasks of performing chest compressions and defibrillation can lead to communication challenges that may prolong time spent off the chest. The purpose of this study was to determine whether defibrillation provided by the provider performing chest compressions led to a decrease in peri-shock pauses as compared to defibrillation administered by a second provider, in a simulated cardiac arrest scenario. METHODS This was a randomized, controlled study measuring pauses in chest compressions for defibrillation in a simulated cardiac arrest model. We approached hospital providers with current CPR certification for participation between July, 2011 and October, 2011. Volunteers were randomized to control (facilitator-administered defibrillation) or experimental (compressor-administered defibrillation) groups. All participants completed one minute of chest compressions on a mannequin in a shockable rhythm prior to administration of defibrillation. We measured and compared pauses for defibrillation in both groups. RESULTS Out of 200 total participants, we analyzed data from 197 defibrillations. Compressor-initiated defibrillation resulted in a significantly lower pre-shock hands-off time (0.57 s; 95% CI: 0.47-0.67) compared to facilitator-initiated defibrillation (1.49 s; 95% CI: 1.35-1.64). Furthermore, compressor-initiated defibrillation resulted in a significantly lower peri-shock hands-off time (2.77 s; 95% CI: 2.58-2.95) compared to facilitator-initiated defibrillation (4.25 s; 95% CI: 4.08-4.43). CONCLUSION Assigning the responsibility for shock delivery to the provider performing compressions encourages continuous compressions throughout the charging period and decreases total time spent off the chest. However, as this was a simulation-based study, clinical implementation is necessary to further evaluate these potential benefits.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Hands-on defibrillation and electrocardiogram artefact filtering technology increases chest compression fraction and decreases peri-shock pause duration in a simulation model of cardiac arrest.

BACKGROUND Reducing pauses during cardiopulmonary resuscitation (CPR) compressions result in better outcomes in cardiac arrest. Artefact filtering technology (AFT) gives rescuers the opportunity to visualize the underlying electrocardiogram (ECG) rhythm during chest compressions, and reduces the pauses that occur before and after delivering a shock. We conducted a simulation study to measure th...

متن کامل

Effects of compression depth and pre-shock pauses predict defibrillation failure during cardiac arrest.

BACKGROUND Cardiopulmonary resuscitation (CPR) and electrical defibrillation are the primary treatment options for ventricular fibrillation (VF). While recent studies have shown that providing CPR prior to defibrillation may improve outcomes, the effects of CPR quality remain unclear. Specifically, the clinical effects of compression depth and pauses in chest compression prior to defibrillation...

متن کامل

Can unannounced simulated cardiac arrest reduce pauses in chest compressions and time to defibrillation?

Background The quality of resuscitation during cardiac arrest in hospital is often suboptimal and there is need for continuous improvement of education. Standardized simulation-based training is proposed as a mean to ensure a consistent and high level of competence among hospital staff, but it is costly and requires frequent recertification. Successful resuscitation is dependent on teamwork and...

متن کامل

A Randomized Controlled Trial of the Effect of Automated External Defibrillator Cardiopulmonary Resuscitation Protocol on Outcome From Out-of-Hospital Cardiac Arrest

Background—Using automated external defibrillators (AEDs) that implement the Guidelines 2000 resuscitation protocol constrains administration of cardiopulmonary resuscitation (CPR) to 50% of AED connection time. We tested a different AED protocol aimed at increasing the CPR administered to patients with out-of-hospital cardiac arrest. Methods and Results—In a randomized controlled trial, patien...

متن کامل

DEFI 2005: a randomized controlled trial of the effect of automated external defibrillator cardiopulmonary resuscitation protocol on outcome from out-of-hospital cardiac arrest.

BACKGROUND Using automated external defibrillators (AEDs) that implement the Guidelines 2000 resuscitation protocol constrains administration of cardiopulmonary resuscitation (CPR) to <50% of AED connection time. We tested a different AED protocol aimed at increasing the CPR administered to patients with out-of-hospital cardiac arrest. METHODS AND RESULTS In a randomized controlled trial, pat...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 15  شماره 

صفحات  -

تاریخ انتشار 2014