Development of a novel low‐radiation‐absorbent lok‐bar to reduce X‐ray scattering and absorption in RapidArc® treatment planning and dose delivery

نویسندگان

  • Hajime Monzen
  • Kazuki Kubo
  • Mikoto Tamura
  • Masaru Hayakawa
  • Yasumasa Nishimura
چکیده

We developed a novel low-radiation-absorbent lok-bar (HM-bar) that is used to secure the immobilizers to the couch. The aim of this study was to investigate the X-ray scattering and absorption properties of the HM-bar in computed tomography (CT) simulation and radiotherapy dose delivery using the Varian Exact™ lok-bar (VL-bar) as a benchmark. CT images were obtained with or without lok-bar, and then each image was visually evaluated for artifacts. The attenuation rates for each lok-bar were measured using a farmer-type ionization chamber (PTW30013) and the I'mRT phantom (IBA Dosimetry GmbH). Measurement points were between gantry angles of 110 and 180°. The treatment apparatus was a NovalisTx (Brainlab AG); X-ray energies were set at 6 MV and 10 MV. In the presence of each lok-bar, the radiation dose was measured in accordance with 10 volumetric modulated arc therapy-stereotactic body radiation therapy (VMAT-SBRT) plans for lung cancer. Artifacts were seldom observed in the CT scans of the HM-bar. The attenuation rate of each lok-bar was higher when the X-ray energy was set at 6 MV than at 10 MV. The highest attenuation rate in the VL-bar was observed at a gantry angle of 112°; the rates were 22.4% at 6 MV and 19.3% at 10 MV. Similarly, the highest attenuation rate for the HM-bar was also observed at a gantry angle of 112°; the rates were 12.2% and 10.1% at 6 MV and 10 MV, respectively. When the VL-bar was evaluated, the isocenter dose of the VMAT-SBRT plans was attenuated by 2.6% as a maximum case. In the case of the HM-bar, the maximum attenuation was 1.4%. In the measurements of each VMAT-SBRT plan, the difference of the dose attenuation rate between the VL-bar and HM-bar was approximately 1%. The HM-bar could be used to minimize the occurrence of artifacts and provide good images in CT scans regarding radiotherapy planning and dose calculation. It can be used for patient therapy at hospitals to provide accurate dose delivery because of its low X-ray scattering and absorption characteristics.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Novel Semisolid Design Based on Bismuth Oxide (Bi2O3) nanoparticles for radiation protection

Objective(S): The dangerous effects of X-ray have been elucidated by scientific studies in occupational health hygiene.  X-ray protective like an apron, thyroid shield and gloves have been made of lead (Pb) to protect against X-ray. However, such equipment makes a lot of safety and health problems such as toxicity, weight, inflexibility and troubles usage in a physician. To ove...

متن کامل

Evaluation of Gold Nanoparticle Size Effect on Dose Enhancement Factor in Megavoltage Beam Radiotherapy Using MAGICA Polymer Gel Dosimeter

Background: Gold nanoparticles (GNPs) are among the most promising radiosensitive materials in radiotherapy. Studying the effective sensitizing factors such as nanoparticle size, concentration, surface features, radiation energy and cell type can help to optimize the effect and possible clinical application of GNPs in radiation therapy. In this study, the radiation sensitive polymer gel was use...

متن کامل

Evaluation of Tumor Control and Normal Tissue Complication Probability in Head and Neck Cancers with Different Sources of Radiation: A Comparative Study

Introduction: The ultimate goal of radiation treatment planning is to yield a high tumor control probability (TCP) with a low normal tissue complication probability (NTCP). Historically  dose volume histogram (DVH) with only volumetric dose distribution was utilized as a popular tool for plan evaluation  hence present study aimed to compare the radiobiological effectiveness of the cobalt-60 (Co...

متن کامل

COVID-19 Tragic Pandemic: Concerns over Unintentional “Directed Accelerated Evolution” of Novel Coronavirus (SARS-CoV-2) and Introducing a Modified Treatment Method for ARDS

Global health authorities are trying to work out the current status of the novel coronavirus (COVID-19) outbreak and explore methods to reduce the rate of its transmission to healthy individuals. In this viewpoint we provide insights concerning how health care professionals can unintentionally shift the novel coronavirus type to more drug-resistant forms. It is worth noting that viruses usually...

متن کامل

Radiosensitization effect of ZnO nanoparticles in lung cancer cells at clinically relevant megavoltage energy

Introduction: Radiation therapy is one of the major modalities that have long been used in cancer treatment. Radiotherapy is often accompanied by early and late toxicity and side effects and narrow therapeutic window. Similarity in radiation absorption properties of tumors and neighboring healthy tissues is often the reason for low specificity of radiation therapy. Development ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2017