Functional differences in bi-level pressure preset ventilators.
نویسندگان
چکیده
The performance of four bilevel positive pressure preset ventilators was compared. The ventilators tested were; BiPAP ST30 (Respironics); Nippy2 (B + D Electrical); Quantum PSV (Healthdyne); and Sullivan VPAP H ST (Resmed). A patient simulator was used to determine the sensitivity of the triggering mechanisms and the responses to a leak within the patient circuit, and to changes in patient effort. Significant differences (p <0.05) between the devices were seen in the trigger delay time and inspiratory trigger pressure. When a leak was introduced into the patient circuit, the fall in tidal volume (VT) was less than ten per cent for each ventilator. The addition of patient effort produced a number of changes in the ventilation delivered. Patient efforts of 0.25 s induced a variable fall in VT. An increase in VT was seen with some ventilators with patient efforts of 1 s but the effect was variable. Trigger failures and subsequent falls in minute volume were seen with the BiPAP and the Nippy2 at the highest respiratory frequency. Differences in the responses of the ventilators are demonstrated that may influence the selection of a ventilator, particularly in the treatment of breathless patients with ventilatory failure.
منابع مشابه
Performance of ventilators for noninvasive positive-pressure ventilation in children.
The aim of the present study was to evaluate the performance characteristics of all the ventilators proposed for home noninvasive positive-pressure ventilation in children in France. The ventilators (one volume-targeted, 12 pressure-targeted and four dual) were evaluated on a bench which simulated six different paediatric ventilatory patterns. For each ventilator, the quality of the inspiratory...
متن کاملHarms of unintentional leaks during volume targeted pressure support ventilation.
BACKGROUND Volume targeted pressure support ventilation (VT-PSV) is a hybrid mode increasingly used to maintain a minimal tidal volume (VT) by automatically adjusting the level of inspiratory pressure. The objective of the study was to determine the ability of home ventilators to maintain the preset minimal VT during unintentional leaks in a VT-PSV mode. METHODS Seven ventilators were tested ...
متن کاملComparison of pressure and volume preset nasal ventilator systems in stable chronic respiratory failure.
Nasal intermittent positive pressure ventilation (NIPPV) has been widely used in the treatment of chronic respiratory disease. Ventilators may be volume or pressure preset; each type has theoretical advantages, but to date there has been no formal comparison. We wanted to assess the efficacy on blood gas changes that may be achieved and overall acceptability of four nasal ventilators (two press...
متن کاملA laboratory comparison of four positive pressure ventilators used in the home.
Four brands of ventilator used for noninvasive positive pressure ventilation in the home were assessed to determine their performance on a patient simulator. We tested the tidal volume (VT) preset Companion 2801 (Puritan Bennett), minute volume preset Monnal D (Taema), and two pressure preset ventilators, the Nippy (Friday Medical) and the Bilevel Positive Airway Pressure (BiPAP) (Respironics)....
متن کاملEvaluation of a fluidic ventilator: a new approach to mechanical ventilation.
T he emerging technology of fluid amplifiers and control devices has provided a new approach to the design and construction of mechanical ventilators. Fluidics, as this technology has come to be known, has not yet found wide application in medicine, and the fundamentals of fluid control devices are not familiar to many in that field. These are basically simple devices that employ fluids (gases ...