Evaluation of a semi-automated blood urea nitrogen analyzer.
نویسندگان
چکیده
We compared results for urea concentrations in plasma and serum as measured with the Kimble Blood Urea Nitrogen Analyzer method as the test method with those determined by continuous-flow analysis (AutoAnalyzer 1 method) as the comparison method. We evaluated accuracy and precision for patients' samples, National Bureau of Standards Standard Reference Material, College of American Pathologists Survey Validated Reference Materials, and College of American Pathologists Quality Assurance Service control materials. We found the test method to be more accurate and precise than the comparison method.
منابع مشابه
Evaluation of a new multichannel analyzer, "Astra-8".
We evaluated a new multiple-channel chemistry analyzer, the Beckman "Astra-8," which is controlled by a microprocessor with 52K bytes of memory. The instrument we tested performed the following tests: Na, K, CO2, chloride, urea nitrogen, glucose, and creatinine. We compared the Astra-8 to a continuous-flow (SMA 6/60) and a discrete (Du Pont aca, urea nitrogen only) analyzer. The Astra-8 demonst...
متن کاملFour commercial urease reagents and a laboratory-prepared reagent compared for analysis of blood urea nitrogen with the Beckman analyzer.
Three laboratory-prepared urease reagents were compared with a commercial preparation supplied for routine use on the Beckman Blood Urea Nitrogen Analyzer. There were discrepancies in results for urea nitrogen among the four urease reagents when matching serum and the corresponding oxalate/fluoride treated plasma were compared as measured with the Beckman Analyzer and continuous-flow (AutoAnaly...
متن کاملExtended clinical trial and evaluation of urea nitrogen determination with the Ektachem GLU/BUN analyzer.
We followed the "abbreviated precision protocol" of the National Committee for Clinical Laboratory Standards for the evaluation of precision, accuracy, and carryover in analysis for urea nitrogen with the multilayer film analysis system ("Ektachem"). We analyzed 456 clinical samples with this instrument, by the manual urease/glucose dehydrogenase method, and with the Beckman System I GLU/BUN An...
متن کاملEvaluation of the effect of exercise on serum creatinine and urea nitrogen in hemodialysis patients
Introduction. Chronic renal failure is a progressive disease in which kidney function is impaired, so that the body can not metabolize and maintain fluid balance and electrolytes. Therefore, the present study investigated changes in serum creatinine and urea nitrogen in hemodialysis patients in response to exercise. Materials and Methods. The subjects of this study were dialysis patients who we...
متن کاملEvaluation of an engineering model of the "EKTACHEM" analyzer for glucose and urea assay.
We report results of a 21-week evaluation of an engineering model of the Kodak “EKTACHEM Glu/BUN Analyzer.” This novel clinical instrument measures glucose and “urea nitrogen” (actually, urea) in a minimum of 30 j.tL of body fluids by use of the principles of thin, multilayer film analysis and reflectance spectrophotometry. We compared the performance of the instrument to that of the Dupont aca...