Characteristics of a novel treatment system for linear accelerator–based stereotactic radiosurgery

نویسندگان

  • Ning Wen
  • Haisen Li
  • Kwang Song
  • Karen Chin‐Snyder
  • Yujiao Qin
  • Jinkoo Kim
  • Maria Bellon
  • Misbah Gulam
  • Stephen Gardner
  • Anthony Doemer
  • Suneetha Devpura
  • James Gordon
  • Indrin Chetty
  • Farzan Siddiqui
  • Munther Ajlouni
  • Robert Pompa
  • Zane Hammoud
  • Michael Simoff
  • Steven Kalkanis
  • Benjamin Movsas
  • M. Salim Siddiqui
چکیده

The purpose of this study is to characterize the dosimetric properties and accuracy of a novel treatment platform (Edge radiosurgery system) for localizing and treating patients with frameless, image-guided stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT). Initial measurements of various components of the system, such as a comprehensive assessment of the dosimetric properties of the flattening filter-free (FFF) beams for both high definition (HD120) MLC and conical cone-based treatment, positioning accuracy and beam attenuation of a six degree of freedom (6DoF) couch, treatment head leakage test, and integrated end-to-end accuracy tests, have been performed. The end-to-end test of the system was performed by CT imaging a phantom and registering hidden targets on the treatment couch to determine the localization accuracy of the optical surface monitoring system (OSMS), cone-beam CT (CBCT), and MV imaging systems, as well as the radiation isocenter targeting accuracy. The deviations between the percent depth-dose curves acquired on the new linac-based system (Edge), and the previously published machine with FFF beams (TrueBeam) beyond D(max) were within 1.0% for both energies. The maximum deviation of output factors between the Edge and TrueBeam was 1.6%. The optimized dosimetric leaf gap values, which were fitted using Eclipse dose calculations and measurements based on representative spine radiosurgery plans, were 0.700 mm and 1.000 mm, respectively. For the conical cones, 6X FFF has sharper penumbra ranging from 1.2-1.8 mm (80%-20%) and 1.9-3.8 mm (90%-10%) relative to 10X FFF, which has 1.2-2.2mm and 2.3-5.1mm, respectively. The relative attenuation measurements of the couch for PA, PA (rails-in), oblique, oblique (rails-out), oblique (rails-in) were: -2.0%, -2.5%, -15.6%, -2.5%, -5.0% for 6X FFF and -1.4%, -1.5%, -12.2%, -2.5%, -5.0% for 10X FFF, respectively, with a slight decrease in attenuation versus field size. The systematic deviation between the OSMS and CBCT was -0.4 ± 0.2 mm, 0.1± 0.3mm, and 0.0 ± 0.1 mm in the vertical, longitudinal, and lateral directions. The mean values and standard deviations of the average deviation and maximum deviation of the daily Winston-Lutz tests over three months are 0.20 ± 0.03 mm and 0.66 ± 0.18 mm, respectively. Initial testing of this novel system demonstrates the technology to be highly accurate and suitable for frameless, linac-based SRS and SBRT treatment.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Development of A Head Docking Device for Linac-based Stereotactic Radiosurgery System to be Installed on Neptun 10 PC Linac

Introduction:  Stereotactic  radiosurgery  is  a  technique  for  the  treatment  of  intracranial  lesions  with  highly  collimated  x-ray  beam.  Linac-based  radiosurgery  is  currently  performed  by  two  techniques:  couch or pedestal mounted. In the first technique, a device is required to affix patient's head to a couch  and  moreover  to  translate  it  accurately.  Structure  of  suc...

متن کامل

Comparison of measured and Monte Carlo calculated dose distributions from “circular collimators” for radiosurgical beams

Background: Stereotactic radiosurgery is an important clinical tool for the treatment of small lesions in the brain, including benign conditions, malignant and localized metastatic tumors. A dosimetry study was performed for Elekta ‘Synergy S ’ as a dedicated Stereotactic radiosurgery unit, capable of generating circular radiation fields with diameters of 1-5 cm at isocentre using the BEAM/EGS4...

متن کامل

Stereotactic radiosurgery for intracranial tumors: a single center experience

Stereotactic radiosurgery is being increasingly used for the treatment of both benign and malignant disorders such as brain metastasis, spinal cord tumors, intracranial blood vessel abnormalities and neurological/functional problems such as Parkinson’s disease, epilepsy, obsessive-compulsive disorder and trigeminal neuralgia. The purpose of this report was to present our experience at the Depar...

متن کامل

Stereotactic Radiosurgery and Stereotactic Body Radiotherapy

Patient Selection Criteria Coverage for the use of stereotactic radiosurgery (SRS) using a gamma or linear-accelerator (LINAC) will be considered when any of the following criteria are met:  Arteriovenous malformations;  Acoustic neuromas;  Pituitary adenomas;  Non-resectable, residual or recurrent meningiomas;  Craniopharyngiomas;  Glomus jugulare tumors;  Solitary or multiple brain met...

متن کامل

Overall and disease-free survival greater than 12 years in metastatic non-small cell lung cancer after linear accelerator-based stereotactic radiosurgery for solitary brain metastasis.

The best treatment approach for solitary brain metastasis is not well defined and there is no consensus on this issue. It is still being debated whether patients with isolated brain metastasis should undergo surgical resection or stereotactic radiosurgery, and which patients should receive adjuvant whole brain radiotherapy. The median survival in patients with single or multiple metastatic lesi...

متن کامل

Stereotactic Radiosurgery in the Treatment of Glioblastoma Multiforme: Current Status of Technology and Potential Role of Microbeam Radiosurgery

Stereotactic radiosurgery is an emerging treatment option offered to patients with glioblastoma (GB). Radiosurgery is performed as an outpatient procedure and provides a safe and effective non-invasive treatment for focal GB. High-energy beams originating from cobalt sources placed into a helmet-shaped device (Gamma Knife®, Elekta) or generated by a linear accelerator (LINAC), rotating on a gan...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015