Pretreatment quality assurance of volumetric modulated arc therapy on patient CT scan using indirect 3D dosimetry system
نویسندگان
چکیده
Purpose: Aim of this study is to clinically implement the COMPASS 3D dosimetry system for pretreatment quality assurance of volumetric modulated arc therapy (VMAT-RapidArc) treatment plans. Methods: For this study, 10 head and neck (H&N) and pelvis VMAT plans dose response from Linac was measured using COMPASS system along with MatriXXEvolution and 3D dose reconstructed in the patient computed tomography (CT) scan. Dose volume histograms and 3D gamma were used to evaluate difference between the measured and calculated values. In order to validate the COMPASS system, dose response for open fields were acquired for both homogeneous and inhomogeneous phantoms. Results: The average dose difference between Eclipse treatment planning system (TPS) calculated and COMPASS measured (homogenous medium) in normalization region, inner region, penumbra region and buildup region was less than ±2%. In inhomogeneous phantom, there was a maximum difference -3.17% in lung, whereas the difference other densities was within ±2%. The systematic increase in the average 3D gamma between the TPS calculated and COMPASS measured for VMAT plans with known dose errors and multi-leaf collimator (MLC) offset errors shows that COMPASS system was sensitive enough to find clinical significant errors. The 3D dose parameters (D95, D1, and average dose) of all H&N and pelvis patients were well within the clinically acceptable tolerance level of ±5%. The average 3D gammas for planning target volumes (PTV) and organ at risks (OAR) of the patients were less than 0.6. Conclusion: The results from this study show that COMPASS along with MatriXXEvolution can be effectively used for pretreatment verification of VMAT plans in the patient anatomy.
منابع مشابه
Performance evaluation of gated volumetric modulated arc therapy
Background: Aim of this study is to evaluate the accuracy of the gated volumetric modulated arc therapy (VMAT/RapidArc) using 2D planar dosimetry, DynaLog files and COMPASS 3D dosimetry system. Materials and Methods: Pre-treatment quality assurance of 10 gated VMAT plans was verified using 2D array and COMPASS 3D dosimetry system. Advantage of COMPASS over 2D planar is that it provides the clin...
متن کاملTrue 3D chemical dosimetry (gels, plastics): Development and clinical role
Since the introduction of volumetric chemical dosimetry with Fricke gel dosimeters in the 1980s, three-dimensional (3D) dosimetry has been a promising technique for the clinic, since it provides a unique methodology for 3D dose measurement of the complex conformal dose distributions achieved by modern techniques such as Intensity Modulated and Volumetric Arc Radiation Therapy. In the last decad...
متن کاملDosimetry for audit and clinical trials: challenges and requirements
Many important dosimetry audit networks for radiotherapy have their roots in clinical trial quality assurance (QA). In both scenarios it is essential to test two issues: does the treatment plan conform with the clinical requirements and is the plan a reasonable representation of what is actually delivered to a patient throughout their course of treatment. Part of a sound quality program would b...
متن کاملQuantitative evaluation of 3D dosimetry for stereotactic volumetric‐modulated arc delivery using COMPASS
The purpose of this study was to evaluate quantitatively the patient-specific 3D dosimetry tool COMPASS with 2D array MatriXX detector for stereotactic volumetric-modulated arc delivery. Twenty-five patients CT images and RT structures from different sites (brain, head & neck, thorax, abdomen, and spine) were taken from CyberKnife Multiplan planning system for this study. All these patients...
متن کاملThe NCS code of practice for the quality assurance and control for volumetric modulated arc therapy
In 2010, the NCS (Netherlands Commission on Radiation Dosimetry) installed a subcommittee to develop guidelines for quality assurance and control for volumetric modulated arc therapy (VMAT) treatments. The report (published in 2015) has been written by Dutch medical physicists and has therefore, inevitably, a Dutch focus. This paper is a condensed version of these guidelines, the full report in...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -