Validation of the RightSpot™ device for determination of gastric pH during nasogastric tube placement

نویسندگان

  • Charles R Lambert
  • David Varlotta
  • Marjorie Posey
  • Jadie L Heberlein
  • Janice M Shirley
چکیده

BACKGROUND The RightSpotpH™ indicator is a new device designed to facilitate pH-based confirmation of nasogastric tube placement while minimizing exposure of the operator to nasogastric contents. STUDY OBJECTIVES In vivo and in vitro validation of the RightSpotpH™ indicator. FINDINGS Patients (23) undergoing general anesthesia had placement of a nasogastric tube and gastric pH electrode catheter following endotracheal intubation. Direct intragastric pH was recorded simultaneously with gastric aspirate pH using the RightSpotpH™ indicator as well as an external pH electrode. In vitro validation of the RightSpotpH™ indicator was performed using standard buffer solutions. Electrode catheter-determined intragastric pH was compared to the RightSpotpH™ indicator result (pH greater or less than 4.5) as was aspirate pH using an external pH electrode. The positive predictive value for the RightSpotpH™ indicator was 0.92 (95% CI 0.62-0.99) and negative predictive value was 1.0 (95% CI 0.48-1.0) when compared to the direct catheter intragastric pH determination. This corresponds to a sensitivity of 1.0 and specificity of 0.83. The positive predictive value for the RightSpotpH™ indicator was 1.0 (95% CI 0.29-0.99) and negative predictive value was 0.75 (95% CI 0.19-0.99) when compared to the external pH electrode determination of aspirate. This corresponds to a sensitivity of 0.91 and specificity of 1.0. Blinded in vitro testing of 46 buffer solution samples of pH 2-7 yielded sensitivity and specificity for correct pH discrimination of 1.0 and 1.0. CONCLUSIONS The RightSpotpH™ indicator is sensitive and specific for determination of intragastric pH determination as less than or greater than 4.5 as commonly used for nasogastric tube placement.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Validation of the RightSpotTM device for determination of gastric pH during nasogastric tube placement

Background: The RightSpotpHTM indicator is a new device designed to facilitate pH-based confirmation of nasogastric tube placement while minimizing exposure of the operator to nasogastric contents. Study objectives: In vivo and in vitro validation of the RightSpotpHTM indicator. Findings: Patients (23) undergoing general anesthesia had placement of a nasogastric tube and gastric pH electrode ca...

متن کامل

Validation of the RightLevelpHTM Detector for Monitoring Gastric pH Running Title: Gastric pH Monitoring

Background: The RightLevelpHTM indicator is a new device designed to measure pH of gastric aspirate obtained via nasogastric tube while minimizing exposure of the operator. Study Objectives: In vivo and in vitro validation of the RightLevelpHTM indicator. Methods: Patients (23) undergoing general anesthesia had placement of a nasogastric tube and gastric pH electrode catheter following endotrac...

متن کامل

Validation of the RightLevelpH Detector for Monitoring Gastric pH.

BACKGROUND The RightLevelpH indicator is a new device designed to measure the pH of gastric aspirate obtained via a nasogastric tube while minimizing exposure of the operator. OBJECTIVE To validate the RightLevelpH indicator in vivo and in vitro. METHODS With general anesthesia, 20 patients had placement of a nasogastric tube and a gastric pH electrode catheter after endotracheal intubation...

متن کامل

Confirming nasogastric tube position: methods & restrictions: A narrative review

Background and Purpose: Inserting a nasogastric tube, though a common clinical procedure with widespread use for critically ill patients, can produce unexpected complications so that tube misplacement into the lungs is a potential complication with serious consequences. The reliability of common bedside methods to differentiate between pulmonary and gastric placement has not been acceptable. Th...

متن کامل

The effectiveness of ultrasonography in verifying the placement of a nasogastric feeding tube in patients with low consciousness at an emergency center

Patients and methods The study included 47 patients who were all over 18 years of age. In all patients, tube placement was verified by chest x-rays. Auscultation, pH analysis of gastric aspirates, and ultrasonography were conducted on each patient in random order. The mean patient age was 57.62 ± 17.24 years, and 28 males (59.6%) and 19 females (40.4%) were included. The nasogastric tube was in...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 6  شماره 

صفحات  -

تاریخ انتشار 2013