Standard Laryngeal Mask Airway and LMA-ProSeal during laparoscopic surgery.
نویسندگان
چکیده
STUDY OBJECTIVE To compare the frequency of airway seal and sore throat with the LMA-ProSeal (PLMA) and the standard Laryngeal Mask Airway (LMA) during laparoscopic surgery. DESIGN Prospective, controlled, randomized, nonblinded clinical study. SETTING University-affiliated hospital. PATIENTS 60 adult, ASA physical status I, II, and III patients undergoing laparoscopic surgery with general anesthesia, without contraindication to the use of the laryngeal mask. INTERVENTIONS Patients were randomized to receive mechanical ventilation [tidal volume (V(T)) 7 mL/kg(-1); positive end-expiratory pressure (PEEP) 10 cmH(2)O] through the PLMA or the standard LMA, both equipped with a gastric tube. MEASUREMENTS Heart rate, arterial pressure, inspiratory and expiratory V(T), airway pressure, end-tidal CO(2) partial pressure, and pulse oximetry were recorded. The leak fraction was calculated as the difference between the inspiratory and expiratory V(T) divided by the inspiratory V(T). Postoperative sore throat frequency was scored in the recovery room ("early") and 1 week after surgery ("late"). MAIN RESULTS All patients were successfully ventilated through the assigned laryngeal mask. The leak fraction was 7 +/- 3% with the LMA and 7 +/- 4% with the PLMA (p = 0.731). In one patient, the PLMA drainage tube was not patent despite a leak fraction of 5%, and there was no clinically detectable air leak. During the recovery room stay, the frequency of sore throat was scored as mild in 13% and 10% of patients with the standard LMA and the PLMA, respectively, and was absent in the remaining patients (p = 0.99, between groups). There were no differences in the frequency of sore throat between the "early" and "late" evaluations (p = 0.99). CONCLUSIONS The PLMA and the LMA show similar airtight efficiency during laparoscopy. The patency of the PLMA drainage tube should always be confirmed. The sore throat evaluation performed in recovery room appears as reliable as later evaluations.
منابع مشابه
Limitation of Laryngeal Mask Airway Usage in Open-Neck Procedure
I read the manuscript entitled “Case Report of An Open Neck Procedure Complication Associated with Laryngeal Mask Airway Use” by Subramaniam T et al. This is a valuable manuscript describing iatrogenic complications during level-2 cervical lymph node biopsy, in which the airway is managed with laryngeal mask airway (LMA). During the procedure, the surgeon realized that the mass was actually the...
متن کاملComparison of Supreme Laryngeal Mask Airway and ProSeal Laryngeal Mask Airway during Cholecystectomy.
OBJECTIVE This study compared the safety and efficacy of the Supreme Laryngeal Mask Airway (S-LMA) with that of the ProSeal-LMA (P-LMA) in laparoscopic cholecystectomy. MATERIAL AND METHODS Sixty adults were randomly allocated. Following anaesthesia induction, experienced LMA users inserted the airway devices. RESULTS Oropharyngeal leak pressure was similar in groups (S-LMA, 27.8±2.9 cmH2O;...
متن کاملComparison of the Laryngeal Mask Airways: Laryngeal Mask Airway-classic and Laryngeal Mask Airway-proseal in Children
Introduction In the past 25 years with the development of various supraglottic airway devices, the armamentarium for airway management has increased. In our study, the laryngeal mask airway (LMA)-ProSeal is compared with the LMA-Classic with respect to as follows: (a) Ease of insertion and number of insertion attempts, (b) Device positional stability and possible oropharyngeal leak, (c) Airway ...
متن کاملCase Report of an open Neck Procedure Complication Associated with Laryngeal Mask Airway Use
Introduction: The laryngeal mask airway (LMA) is a safe method to establish airway control during general anaesthesia. In otolaryngology surgery, the use of a LMA is well established in ear surgery; however, the use of a LMA during open neck procedures remains controversial. We report a case in which the pharyngeal distortion by the LMA cuff resulted in an iatrogenic complication. Case Report...
متن کاملA Comparative Study of ProSeal Laryngeal Mask Airway Cuff Pressure Changes with and without use of Nitrous oxide in Laparoscopic Surgeries
ProSeal Laryngeal mask airway is increasingly being used as an airway device for laparoscopic surgery. Nitrous oxide can diffuse into the cuff of airway devices and may further increase the intracuff pressure. The present study was designed to investigate the intracufff pressure changes during anesthesia with and without use of nitrous oxide in patients undergoing laparoscopic surgery and also ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of clinical anesthesia
دوره 15 6 شماره
صفحات -
تاریخ انتشار 2003