Totally implantable venous access devices in cystic fibrosis: complications and patients' opinions.

نویسندگان

  • H C Rodgers
  • K Liddle
  • S J Nixon
  • J A Innes
  • A P Greening
چکیده

The introduction of totally implantable venous access devices (TIVAD) has provided a solution to difficult venous access in patients with cystic fibrosis. Early reports have, however, recognized a number of complications with their use. We report our experience with five devices used over 8 yrs with regard to complications and patient attitudes. Patients' notes were reviewed to record the details of TIVAD insertion, duration of function, and complications. In January 1996 the surviving 30 patients were surveyed on their attitudes to TIVAD and complications by written questionnaire. Sixty one ports were implanted in 42 patients (aged 16-47 yrs) between June 1988 and January 1996, giving a total of 1,510 patient-months' experience. The duration of function ranged from 2 weeks to 6 yrs. Survival analysis showed that the median survival of ports was 53 months, 42 out of 61 (69%) had not failed in service at the end of follow-up or patient death. Twenty-three complications occurred in 19 patients. These included: line occlusion (10 patients), venous thrombosis (4), difficult access (3), infection (2), cellulitis (1), inversion of port chamber (2) and pneumothorax (1). The questionnaire showed that patients had strong views on the positioning of their port. Lifestyle issues included interference with seatbelts (8 patients), sport (4), clothing (2), sexual relations (2) and cosmetic appearance (15). Complication rates were similar to those in other studies, although infection rates and salvage of an occluded port were lower. The survey highlighted a number of lifestyle issues, with cosmetic appearance deemed unsatisfactory by half of the patients. However, the majority (28 out of 30) believed their totally implantable venous access devices to be a better alternative to cannulae or long lines.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Totally implantable venous access devices in children with cystic fibrosis: incidence and type of complications.

BACKGROUND Totally implantable vascular access devices (TIVADs) are accepted as a safe and effective method of facilitating long term intravenous therapy. We report our experience of the use of these devices in children with cystic fibrosis with a particular focus on the incidence and type of complications. METHODS The medical records of patients with cystic fibrosis who underwent placement o...

متن کامل

Implantable Port Devices, Complications and outcome in Pediatric Cancer, a Retrospective Study

Abstract Background Peripheral blood vessels, due to availability are used for many years in cancer patients, however in patients with potentially harmful drugs to skin (vesicant drugs) or difficult accessibility to vessels, the use of implantable port (totally implantable venous access port-TIVAP)devices with central vascular access are important.  ...

متن کامل

Follow-up of 452 totally implantable vascular devices in cystic fibrosis patients.

The use and complications of totally implantable vascular access devices (TIVADs) were examined during multiple courses of antibiotics in cystic fibrosis (CF) patients. This retrospective study involved 36 CF centres. Risk factors for removal and septicaemia were sought by survival analysis of censored data. Multivariate Cox models were constructed with removal or septicaemia as the event and t...

متن کامل

Effectiveness of Permanent Implantable Catheter (Polysite) in Children with Cancer

Abstract Background Totally implantable central venous access devices (ports) have been available for over 10 years, but have not been achieved widespread use in pediatric oncology patients. Ports facilitate the administration of chemotherapy in children with cancer. Materials and Methods In this study, early complications of implantable central venous access devices in children with diff...