Failure mode and effects analysis of skin electronic brachytherapy using Esteya® unit

نویسندگان

  • Blanca Ibanez-Rosello
  • Juan Antonio Bautista-Ballesteros
  • Jorge Bonaque
  • Francisco Celada
  • Françoise Lliso
  • Vicente Carmona
  • Jose Gimeno-Olmos
  • Zoubir Ouhib
  • Joan Rosello
  • Jose Perez-Calatayud
چکیده

PURPOSE A failure mode and effect analysis (FMEA) of skin lesions treatment process using Esteya™ device (Elekta Brachyterapy, Veenendaal, The Netherlands) was performed, with the aim of increasing the quality of the treatment and reducing the likelihood of unwanted events. METHODS A multidisciplinary team with experience in the treatment process met to establish the process map, which outlines the flow of various stages for such patients undergoing skin treatment. Potential failure modes (FM) were identified and the value of severity (S), frequency of occurrence (O), and lack of detectability (D) of the proposed FM were scored individually, each on a scale of 1 to 10 following TG-100 guidelines of the AAPM. These failure modes were ranked according to our risk priority number (RPN) and S scores. The efficiency of existing quality management tools was analyzed through a reassessment of the O and D made by consensus. RESULTS 149 FM were identified, 43 of which had RPN ≥ 100 and 30 had S ≥ 7. After introduction of the tools of quality management, only 3 FM had RPN ≥ 100 and 22 FM had RPN ≥ 50. These 22 FM were thoroughly analyzed and new tools for quality management were proposed. The most common cause of highest RPN FM was associated with the heavy patient workload and the continuous and accurate applicator-patient skin contact during the treatment. To overcome this second item, a regular quality control and setup review by a second individual before each treatment session was proposed. CONCLUSION FMEA revealed some of the FM potentials that were not predicted during the initial implementation of the quality management tools. This exercise was useful in identifying the need of periodic update of the FMEA process as new potential failures can be identified.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetric characteristics of a new unit for electronic skin brachytherapy

PURPOSE Brachytherapy with radioactive high dose rate (HDR) (192)Ir source is applied to small skin cancer lesions, using surface applicators, i.e. Leipzig or Valencia type. New developments in the field of radiotherapy for skin cancer include electronic brachytherapy. This technique involves the placement of an HDR X-ray source close to the skin, therefore combining the benefits of brachythera...

متن کامل

Clinical implementation of a new electronic brachytherapy system for skin brachytherapy

Although surgery is usually the first-line treatment for nonmelanoma skin cancers, radiotherapy (RT) may be indicated in selected cases. Radiation therapy as primary therapy can result in excellent control rates, cosmetics, and quality of life. Brachytherapy is a radiation treatment modality that offers the most conformal option to patients. A new modality for skin brachytherapy is electronic b...

متن کامل

Commissioning and periodic tests of the Esteya® electronic brachytherapy system

A new electronic brachytherapy unit from Elekta, called Esteya(®), has recently been introduced to the market. As a part of the standards in radiation oncology, an acceptance testing and commissioning must be performed prior to treatment of the first patient. In addition, a quality assurance program should be implemented. A complete commissioning and periodic testing of the Esteya(®) device usi...

متن کامل

Efficacy and safety of electronic brachytherapy for superficial and nodular basal cell carcinoma

PURPOSE Surface electronic brachytherapy (EBT) is an alternative radiotherapy solution to external beam electron radiotherapy and high-dose-rate radionuclide-based brachytherapy. In fact, it is also an alternative solution to surgery for a subgroup of patients. The objective of this work is to confirm the clinical efficacy, toxicity and cosmesis of a new EBT system, namely Esteya(®) in the trea...

متن کامل

Independent assessment of source transit time for the BEBIG SagiNova® high dose rate brachytherapy afterloader

Introduction: The dwell time and transit time components contribute to the overall delivered dose to patients in high dose rate (HDR) brachytherapy treatments. The transit time results from source entry and exit as well as source movements between dwell positions. It depends on various parameters such as the source speed profile, source indexer distance to dwell position, and ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 8  شماره 

صفحات  -

تاریخ انتشار 2016