Diagnostic accuracy of the iCare rebound tonometer compared to the Perkins applanation tonometer in assessing intraocular pressure in rural patients.

نویسندگان

  • Yifan Li
  • Christopher R Carpenter
  • Kathryn Nicholson
  • William Ken Milne
چکیده

BACKGROUND Vision health is recognized as a critical unmet need in North America. The ocular morbidity associated with glaucoma results from increased intraocular pressure (IOP) and early detection is crucial for the management of glaucoma. Our objective was to find a diagnostically accurate screening tool for intraocular hypertension that can be used in rural communities. We sought to validate the diagnostic accuracy of the iCare rebound tonometer against the gold standard Perkins applanation tonometer (PAT) in measuring IOP. METHODS Patients from two rural communities in Ontario, Canada visiting their optometrists for routine appointments had their IOP measured by a non-contact tonometer (NCT), an iCare rebound tonometer, and a Perkins applanation tonometer (PAT). Values of sensitivity, specificity, and likelihood ratios for a positive and negative result were calculated for the iCare and the NCT. RESULTS Complete data was collected from 209 patients. Overall, the iCare tonometer had high levels of validity, as compared to the gold standard PAT. The iCare tonometer displayed excellent sensitivity of 98.3% (90-99%, 95% CI) and excellent negative likelihood ratio of 0.024 (0.0088-0.066, 95% CI) which is useful for ruling out intraocular hypertension. CONCLUSIONS The iCare tonometer is a reasonably valid tool for detecting elevated IOP. Its ease of use, simplicity, and accessibility makes it a good screening tool to improve eye health in rural areas.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

BACKGROUND To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients. DESIGN A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic. PARTICIPANTS OR SAMPLES One hundred nine (54 males : 55 females) patients includ...

متن کامل

Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device.

PURPOSE (a) To investigate whether the ICare rebound tonometer can provide accurate measurements of intraocular pressure (IOP) in the hands of an inexperienced user compared with ICare measurements and Goldmann tonometry by a trained technician and (b) to assess the intrauser reproducibility of IOP measurements and the learning curve among patients using the ICare rebound tonometer. METHODS A...

متن کامل

Accuracy of the ICare rebound tonometer in glaucomatous eyes with topical ocular hypotensive medication.

BACKGROUND The purpose of this study was to assess the accuracy of the ICare tonometer, using the Perkins applanation tonometer (AT) as the reference, in a sample population being treated with travaprost 0.004% for glaucoma. MATERIAL AND METHODS Twenty-eight consecutive patients with open-angle glaucoma or ocular hypertension who had been receiving travaprost 0.004% to control elevated intrao...

متن کامل

Evaluation of a new rebound tonometer for self-measurement of intraocular pressure Sub-title Self-measurement using Icare Home showed good agreement with Goldmann applanation

Title Evaluation of a new rebound tonometer for self-measurement of intraocular pressure Subtitle Self-measurement using Icare Home showed good agreement with Goldmann applanation tonometry. Icare Home self-measurement was well accepted and reliable results obtained by 3 in 4 subjects, suggesting potential use for home measurement of intraocular pressure. Number of references = 26 Number of sup...

متن کامل

Does Rebound Tonometry Probe Misalignment Modify Intraocular Pressure Measurements in Human Eyes?

Purpose. To examine the influence of positional misalignments on intraocular pressure (IOP) measurement with a rebound tonometer. Methods. Using the iCare rebound tonometer, IOP readings were taken from the right eye of 36 healthy subjects at the central corneal apex (CC) and compared to IOP measures using the Goldmann applanation tonometer (GAT). Using a bespoke rig, iCare IOP readings were al...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Diagnosis

دوره 2 4  شماره 

صفحات  -

تاریخ انتشار 2015