SNR Quantification with Phased-Array Coils and Parallel Imaging for 3D-FSE

نویسندگان

  • C. Q. Li
  • W. Chen
  • P. J. Beatty
  • A. C. Brau
  • B. A. Hargreaves
  • R. F. Busse
  • G. E. Gold
چکیده

INTRODUCTION: Recently a novel 3D fast-spin-echo sequence (3D-FSE-Cube) has been developed for structural imaging of the head, abdomen and extremities [1,2]. The application of parallel imaging, phased-array coils and partial Fourier acquisition alters the spatial and statistical distributions of image noise, respectively [3]. This study measured noise via noise-only acquisitions (acquired with no RF excitation) and compared resulting signal-to-noise and contrast-to-noise ratio (SNR and CNR) measurements with the traditional foreground and background volume of interest (VOI) estimation method [3,4]. To measure noise statistics in noise-only acquisitions, noise should be processed through the identical linear reconstruction pipeline as signal data, which may involve data-dependent steps such as parallel imaging, partial Fourier homodyne and multichannel reconstructions [5]. METHODS: The right knees of 7 healthy volunteers (mean age 27 yrs, range 21–32 yrs) were imaged using a GE Signa 3T MRI and an 8-channel transmit-receive knee coil. Each subject was imaged 20–22 times with sagittal proton-density 3D-FSE-Cube with different TR, BW, ETL, NEX and acceleration factor combinations for a total of 146 scans. Imaging parameter ranges used were TR 1000–5750ms, BW ±31.25–125kHz, ETL 45–120, NEX 0.5–2 and autocalibrated parallel imaging acceleration factors of 1–3.75. All scans utilized fat saturation and the following fixed acquisition parameters: 35ms TEeff, 256x256 matrix, 16cm FOV, 180 slices at 0.6mm slice thickness. Noise-only data was acquired with RF turned off for each unique combination of parameters. Both signal and noise-only data were reconstructed offline to apply parallel imaging calibration weights, homodyne partial Fourier phase correction and multichannel image combination to the noise-only data (see Figure 1). For all 146 scans, SNRs were calculated for muscle, cartilage and synovial fluid, and CNRs were calculated for fluid and cartilage. VOIs were placed in the reconstructed signal images to measure mean signal intensities and duplicated in the noise-only reconstructions to measure noise. SNR and CNR were also calculated using the traditional foreground and background method. Tissue VOIs were maintained from the prior SNR measurement while noise was quantified as the standard deviation of background noise in a signal-free region anterolateral to the knee. Traditional SNR and CNR measurements using the foreground and background approach were compared to SNR and CNR calculated from signal and noise-only acquisitions using paired t-tests. RESULTS: SNRs calculated for muscle, cartilage and synovial fluid were significantly overestimated using the traditional foreground and background method (p < 0.0001). Average errors relative to measurements from the noise-only acquisitions were 299%, 376%, 224% and 250% for fluid SNR, muscle SNR, cartilage SNR and fluid-cartilage CNR, respectively (see Figure 2). Because signal intensity was unchanged for all SNR and CNR measurements, overestimation of SNR and CNR is attributable to underestimation of image noise. This observation follows the theoretical prediction of reduced noise in regions without signal due to spatially-varying geometry factors in parallel imaging [6]. CONCLUSION: Acquisition and reconstruction of noise-only data parallel to signal data provides measurements of image SNR and CNR significantly different from traditional foreground and background estimation methods in 3D-FSE-Cube acquisitions. While the reconstruction pipeline described was tailored for 3D-FSE-Cube acquisitions, this noise measurement technique is generalizable to any sequences utilizing parallel imaging, partial Fourier acquisition and multichannel image combination.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Improved high resolution imaging with 4-element liquid nitrogen phased array coil and VD-AUTO-SMASH at 1.5T

W. Kwok, Z. You, J. Zhong University of Rochester, Rochester, NY, United States Synopsis We have recently developed liquid-nitrogen (LN2) cooled phased array RF coils that improve SNR significantly over similar room-temperature array coils and provide larger field-of-view coverage over single coils. In this study, we investigate the feasibility of combining LN2 phased array coils with parallel ...

متن کامل

Phased-array MRI of canine prostate using endorectal and endourethral coils.

A four-channel phased array consisting of one surface coil, two endorectal coils, and one flexible endourethral loop coil was designed for MRI of the canine prostate. The endorectal coils provide high signal in the posterior region of the prostate, while the endourethral and surface coils are sensitive to the central and anterior regions of the prostate. Gel phantom experiments indicate that th...

متن کامل

Parallel imaging of knee cartilage at 3 Tesla.

PURPOSE To evaluate the feasibility and reproducibility of quantitative cartilage imaging with parallel imaging at 3T and to determine the impact of the acceleration factor (AF) on morphological and relaxation measurements. MATERIALS AND METHODS An eight-channel phased-array knee coil was employed for conventional and parallel imaging on a 3T scanner. The imaging protocol consisted of a T2-we...

متن کامل

A 4-channel 3 Tesla phased array receive coil for awake rhesus monkey fMRI and diffusion MRI experiments.

Awake monkey fMRI and diffusion MRI combined with conventional neuroscience techniques has the potential to study the structural and functional neural network. The majority of monkey fMRI and diffusion MRI experiments are performed with single coils which suffer from severe EPI distortions which limit resolution. By constructing phased array coils for monkey MRI studies, gains in SNR and anatom...

متن کامل

A Two Element Phased Array Coil Enabling Widespread Application of High Resolution MR Coronary Angiography

Parallel imaging using sophisticated receiver coils has improved the clinical feasibility of magnetic resonance coronary angiography (MRCA). These techniques, however, are not readily available outside advanced imaging centers. Our custom-made 2-element phased array coil is readily and inexpensively assembled to address this limitation and enable the widespread application of MRCA. The 2-elemen...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2009