Evaluation of effect of specimen-handling parameters for plasma preparation tubes on viral load measurements obtained by using the Abbott RealTime HIV-1 load assay.

نویسندگان

  • Helen Fernandes
  • Svetlana Morosyuk
  • Klara Abravaya
  • Madhuri Ramanathan
  • Lynne Rainen
چکیده

HIV-1 viral load testing is essential to the management of HIV-1-infected patients, and proper specimen handling ensures accurate viral load (VL) results. This study was performed to (i) evaluate the effect of freezing plasma in situ in BD Vacutainer plasma preparation tubes (PPT) on the accuracy of HIV-1 viral load results using the Abbott RealTime HIV-1 assay and (ii) evaluate the effect of whole-blood storage in the PPT for 6 h at room temperature prior to centrifugation (PPT6H) rather than 2 h as specified in the PPT product insert. Of the 64 HIV-positive subjects evaluated, 29 had average viral load counts of >40 copies/ml in at least one of the tubes tested and 35 subjects had a result of either "undetected target" or "below the limit of quantification" (LOQ) for all or some of the tubes regardless of handling condition. For the 29 subjects with VLs that were >LOQ, the mean biases between plasma from Vacutainer K(2)EDTA tubes and plasma frozen in situ in PPT and between K(2)EDTA tube plasma and plasma from PPT6H (log(10) copies/ml) were 0.005 and -0.001, respectively, and r(2) was >0.92 for all correlations. We conclude that VLs determined from plasma frozen in situ in PPT are equivalent to VLs in K(2)EDTA tube plasma and can be used for accurate quantification of HIV-1 RNA in the Abbott RealTime HIV-1 assay. Furthermore specimens collected in PPT can be stored for 6 h at room temperature with no effect on viral load results as measured by the Abbott RealTime HIV-1 assay.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance evaluation of the COBAS/TaqMan HIV-1 v2.0 in HIV-1 positive patients with low viral load: a comparative study.

HIV-1 viral load determination is a crucial step for monitoring the efficacy of highly active antiretroviral therapy (HAART) and predicts disease progression. Real-time PCR based assays are available for monitoring the viral load. They differ in sensitivity, genomic target region and dynamic range. In this study, the performance of the Roche Cobas Taqman HIV-1 v2.0 was evaluated on plasma sampl...

متن کامل

Stability of plasma human immunodeficiency virus load in VACUTAINER PPT plasma preparation tubes during overnight shipment.

VACUTAINER PPT plasma preparation tubes were evaluated to determine the effects of various handling and shipping conditions on plasma human immunodeficiency virus (HIV) load determinations. Plasmas obtained from PPT tubes stored and shipped under nine different conditions were compared to conventional EDTA tube plasmas stored at -70 degrees C within 2 h after phlebotomy. Compared to viral loads...

متن کامل

A RealTime HIV-1 viral load assay for automated quantitation of HIV-1 RNA in genetically diverse group M subtypes A-H, group O and group N samples.

The Abbott RealTime HIV-1 assay is an automated test for monitoring HIV-1 viral load in plasma samples. The assay uses reverse transcription polymerase chain reaction (RT-PCR) technology with homogeneous real-time fluorescent detection. Automated sample preparation is performed on the m2000sp instrument where RNA is isolated using magnetic microparticle technology and dispensed to a PCR tray to...

متن کامل

Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays.

The implementation of antiretroviral therapy demands the need for increased access to viral load (VL) monitoring. Newer real-time VL testing technologies are faster and have larger dynamic ranges and fully automated extraction to benefit higher throughputs in resource-poor environments. The Abbott RealTime human immunodeficiency virus type 1 (HIV-1) assay was evaluated as a new option for testi...

متن کامل

Performance evaluation of the QIAGEN EZ1 DSP Virus Kit with Abbott RealTime HIV-1, HBV and HCV assays.

BACKGROUND Automated sample preparation systems must meet the demands of routine diagnostics laboratories with regard to performance characteristics and compatibility with downstream assays. OBJECTIVES In this study, the performance of QIAGEN EZ1 DSP Virus Kit on the BioRobot EZ1 DSP was evaluated in combination with the Abbott RealTime HIV-1, HCV, and HBV assays, followed by thermalcycling a...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 48 7  شماره 

صفحات  -

تاریخ انتشار 2010