Doubled Scan Efficiency in Multi-slice Real-time Cine TrueFISP Cardiac Imaging by Pseudo-Continuous Sliding Slice Acquisition

نویسندگان

  • A. Greiser
  • S. Zuehlsdorff
  • E. Mueller
چکیده

Introduction: Real-time TrueFisp cardiac imaging allows multiple slice cine acquisitions within a single breathhold. Typically, the steady state preparation is established utilizing a full heartbeat per slice. Hence, only every other heartbeat is used to collect data. A simple elimination of the preparation heartbeats to increase scan efficiency compromises the contrast between blood and myocardium and signal dynamics for the first few cardiac phases. This work presents a novel method that shifts the slice position between consecutive phases by a distance small compared to slice thickness. High contrast is maintained throughout all cardiac phases providing whole heart coverage within a single breathhold. This approach was compared to to a clinically established real-time cine protocol with and without dummy heartbeats for steady state preparation

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Myocardial perfusion imaging using CAIPIRINHA accelerated multi-slice saturation recovery TrueFISP

Introduction: Saturation recovery (SR) prepared TrueFISP sequences are widely in use for myocardial perfusion imaging because of a high intrinsic signal-to-noise ratio (SNR). Since the image acquisition is limited by the heart cycle, a complete coverage of the heart with consecutively acquired slices is difficult. Parallel imaging techniques are required significantly reducing the acquisition t...

متن کامل

CAIPIRINHA accelerated simultaneous multi-slice TrueFISP real-time imaging

Introduction: CAIPIRINHA multi-slice imaging offers the possibility to acquire several imaging slices simultaneously while maintaining the image quality (1), allowing a high acceleration of the imaging procedure. By providing the individual slices with different RF-phase cycles, the simultaneously excited slices are shifted with respect to each other in the FOV, facilitating the separation of t...

متن کامل

Temporal interpolation of real-time cine images for ventricular function assessment

Background In uncooperative patients or in presence of arrhythmia, real-time free-breathing cine MR approaches may offer an alternative to well-established segmented data acquisition strategies [1,2]. Segmented acquisition strategies collect data throughout a number of heartbeats and typically a fixed number of cardiac phases are reconstructed for each slice representing an average RR interval....

متن کامل

Initial experience with a cardiac multi-contrast real-time cine prototype integrating sparse sampling and iterative reconstruction

Methods All patients were submitted to a conventional cardiac magnetic resonance study (Magnetom Aera, Siemens AG Healthcare, Germany) that included shortand long-axis steady-state free-precession (SSFP) segmented cine measurements (spatial resol.: 1.5x1.5 mm; slice thickness: 7 mm; temporal resol.: 40 ms; 7 heart beats (HB)/slice), modified Look-Locker inversion recovery post-contrast T1 mappi...

متن کامل

Accelerated Simultaneous Multi-slice Cardiac Cine Imaging Using a Combination of CAIPIRINHA and Compressed Sensing

images of the two simultaneously excited slices. Displayed are (a) the conventional reconstruction of the CAIPIRINHA acquisition using GRAPPA and (b) the reconstruction of the incoherently undersampled CAIPIRINHA acquisition using the proposed combination of Compressed Sensing and GRAPPA. (c) Difference between a) and b) magnified by a factor of 10. Accelerated Simultaneous Multi-slice Cardiac ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2008