Validation of a sensitivity performance index test protocol and evaluation of colour Doppler sensitivity for a range of ultrasound scanners.

نویسندگان

  • Jacinta E Browne
  • Amanda J Watson
  • Peter R Hoskins
  • Alex T Elliott
چکیده

The ability to detect flow is the most crucial aspect of an ultrasound (US) system because, if flow cannot be detected, no other aspect of performance matters. The objectives of this study were to validate a Doppler "sensitivity performance index," a figure of merit, and to determine if it could be used to differentiate colour Doppler sensitivity performance in scanners of varying complexity. The sensitivity performance index was developed to give a combined measure of related aspects of sensitivity, such as the lowest detectable velocity, the vessel size and the penetration depth. The colour Doppler sensitivity was evaluated objectively as the lowest detectable velocity signal from the deepest achievable point within the Doppler sensitivity phantom free from extraneous noise in a small diameter vessel (3.2 mm inner diameter). The effect of vessel size and mean velocity on the sensitivity performance index were investigated and it was found that the index was not proportional to vessel size, but this may be accounted for by considering the effect of the acoustic properties of the vessel material, the clutter filter and beam shape. The results obtained using flow phantoms with vessel sizes different from those used in this study are, therefore, not directly comparable to the results found in this study; however, a similar trend should be found in the results for the effect of control settings and a similar range of US scanners. It was found that the Doppler sensitivity performance index was a robust challenging test because none of the US scanners evaluated was capable of achieving the highest sensitivity performance index score, which would be limited by the lowest pump velocity and the deepest point of the vessel within the flow phantom. Therefore, this suggests that this method of determining Doppler sensitivity performance is valuable in the absence of other suitable methods, despite the fact that the relationship between the sensitivity performance index and vessel size is not proportional. Furthermore, use of the Doppler sensitivity performance index for the evaluation of a range of scanners demonstrated that curvilinear transducers have higher sensitivity performance indices than higher-frequency linear transducers, due to the higher achievable penetration depth. The effect of instrument settings was assessed for two transducers, the 4C3 curvilinear general-purpose transducer (Aspen) and the PVM375AT curvilinear general-purpose transducer (Nemio). The colour Doppler sensitivity performance was found to be significantly dependent on the clutter filter setting and the output power setting for both transducers tested. Users need to be aware of the effect of these settings on the colour Doppler sensitivity performance of their US scanner when interpreting the clinical significance of the colour Doppler information.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A new ultrasound sign in the diagnosis of pediatric maxillary sinusitis

      Background: Computed tomography scans (CT scan) and X-rays are used to diagnose paediatric maxillary sinusitis. This study aimed at exploring the diagnostic value of the conventional and colour Doppler ultrasounds and their specific findings in cases of paediatric sinusitis.    Methods: A total of 60 children diagnosed with sinusitis were included in this s...

متن کامل

Performance comparison of four commercial GE discovery PET/CT scanners: A monte carlo study using GATE

  Combined PET/CT scanners now play a major role in medicine for in vivo imaging in oncology, cardiology, neurology, and psychiatry. As the performance of a scanner depends not only on the scintillating material but also on the scanner design, with regards to the advent of newer scanners, there is a need to optimize acquisition protocols as well as to compare scanner ...

متن کامل

A review of Doppler ultrasound quality assurance protocols and test devices.

In this paper, an overview of Doppler ultrasound quality assurance (QA) testing will be presented in three sections. The first section will review the different Doppler ultrasound parameters recommended by professional bodies for use in QA protocols. The second section will include an evaluation and critique of the main test devices used to assess Doppler performance, while the final section of...

متن کامل

The Role of Uterine Artery Doppler Ultrasound in the Second Trimester in Predicting Preeclampsia

Background Preeclampsia is a specific pregnancy syndrome. Early diagnosis and proper administration of preeclampsia can lead to improved maternal and neonatal outcome. We aimed to investigate the uterine artery Doppler ultrasound in predicting the preeclampsia in women with a high-risk pregnancy. Materials and Methods This analytic study was conducted on 160 primigravida women with gestational...

متن کامل

Evaluation of Soil Liquefaction Potential by Sensitivity Analysis, Reliability and Data validation

In this study, it is attempted to analyze sensitivity and reliability in order to evaluate the liquefaction potential in soil layers in Tabriz. 62 boreholes that had possible conditions for liquefaction were selected. Seismic mapping was simulated using finite fault method and then the effect of soil layers on PGA was estimated. In continue, the liquefaction potential index was estimated and th...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Ultrasound in medicine & biology

دوره 30 11  شماره 

صفحات  -

تاریخ انتشار 2004