Caval collapse during cardiopulmonary bypass: a reproducible bench model.
نویسندگان
چکیده
OBJECTIVES During open heart surgery, so-called atrial chatter, a phenomenon due to right atria and/or caval collapse, is frequently observed. Collapse of the cava axis during cardiopulmonary bypass (CPB) limits venous drainage and may result downstream in reduced pump flow on (lack of volume) and upstream in increased after-load (stagnation), which in turn may both result in reduced or even inadequate end-organ perfusion. The goal of this study was to reproduce venous collapse in the flow bench. METHODS In accordance with literature for venous anatomy, a caval tree system is designed (polyethylene, thickness 0.061 mm), which receives venous inflow from nine afferent veins. With water as medium and a preload of 4.4 mmHg, the system has an outflow of 4500 ml/min (Scenario A). After the insertion of a percutaneous venous cannula (23-Fr), the venous model is continuously served by the afferent branches in a venous test bench and venous drainage is augmented with a centrifugal pump (Scenario B). RESULTS With gravity drainage (siphon: A), spontaneously reversible atrial chatter can be generated in reproducible fashion. Slight reduction in the outflow diameter allows for generation of continuous flow. With augmentation (B), irreversible collapse of the artificial vena cava occurs in reproducible fashion at a given pump speed of 2300 ± 50 RPM and a pump inlet pressure of -112 mmHg. Furthermore, bubbles form at the cannula tip despite the fact that the entire system is immersed in water and air from the environment cannot enter the system. This phenomenon is also known as cavitation and should be avoided because of local damage of both formed blood elements and endothelium, as well embolization. CONCLUSIONS This caval model provides a realistic picture for the limitations of flow due to spontaneously reversible atrial chatter vs irreversible venous collapse for a given negative pressure during CPB. Temporary interruption of negative pressure in the venous line can allow for recovery of venous drainage. This know-how can be used not only for testing different cannula designs, but also for further optimizing perfusion strategies.
منابع مشابه
Venous cannula performance assessment in a realistic caval tree model.
OBJECTIVES A new caval tree system was designed for realistic in vitro simulation. The objective of our study was to assess cannula performance for virtually wall-less versus standard percutaneous thin-walled venous cannulas in a setting of venous collapse in case of negative pressure. METHODS For a collapsible caval model, a very flexible plastic material was selected, and a model with nine ...
متن کاملThe effects of venous cannulation technique and cardioplegia type on plasma potassium concentration and arterial blood pressure during cardiopulmonary bypass.
The cannulation method and cardioplegia solution used during cardiopulmonary bypass (CPB) may both influence plasma potassium concentrations ([K+]) and mean arterial blood pressure (MAP). Bi-caval or right atrial cannulation methods are routinely used in conjunction with crystalloid or blood cardioplegia. We investigated the influence of cannulation method and cardioplegia solutions on plasma [...
متن کاملHow to improve flow during cardiopulmonary bypass in an acardia experimental model.
OBJECTIVES In extreme scenarios, such as hyperacute rejection of heart transplant, an urgent heart explantation might be necessary. The aim of this experimental study was to determine the feasibility and to improve the haemodynamics of a venoarterial cardiopulmonary bypass after cardiectomy. METHODS A venoarterial cardiopulmonary bypass was established in seven calves (56.4 ± 7 kg) by the trans...
متن کاملRenal angiomyolipoma with inferior vena caval involvement.
Renal angiomyolipomas are innocuous benign tumours which rarely behave aggressively. This is a case of a 48 year old Malay lady presenting with right sided abdominal pain associated with a large right sided abdominal mass. She was diagnosed with renal angiomyolipoma of the right kidney complicated by inferior vena caval tumour thrombosis. She successfully underwent a radical nephrectomy and inf...
متن کاملMitral valve repair with decalcification of the annulus and pericardial patch repair via the trans-septal approach.
Median sternotomy was performed and cardiopulmonary bypass was established using bi-caval cannulation. Cardiac standstill was obtained with antegrade cardioplegia after cross-clamping the aorta. The caval tapes were snared and the right atrium (RA) was opened. An incision was made in the fossa ovalis and was extended into the left atrium (LA). Stay sutures were placed on the edge of the incisio...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery
دوره 46 2 شماره
صفحات -
تاریخ انتشار 2014