Spatial variation of Dosimetric Leaf Gap and the impact in Dose delivery Our work evaluates the variation of Dosimetric Leaf Gap (DLG) along the travel path of each

ثبت نشده
چکیده

Our work evaluates the variation of Dosimetric Leaf Gap (DLG) along the travel path of each Varian Millennium 120-MLC leaf pairs. Traditionally, the Treatment Planning Systems (TPS) utilize a single value of DLG measured under the middle leaf-pair. The TPS models the shape of the leaf tip as straight as opposed to rounded edge. To overcome this limitation, the TPS retracts each leaf tip by half of the value of the DLG so the gap between a fully closed leaf pair equals the DLG parameter.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of fluence‐based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG)

The Eclipse treatment planning system uses a single dosimetric leaf gap (DLG) value to retract all multileaf collimator leaf positions during dose calculation to model the rounded leaf ends. This study evaluates the dosimetric impact of the 2D variation of DLG on clinical treatment plans based on their degree of fluence modulation. In-house software was developed to retrospectively apply the 2D...

متن کامل

Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D

Background: Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accur...

متن کامل

The dosimetric impact of control point spacing for sliding gap MLC fields

Dynamic sliding gap multileaf collimator (MLC) fields are used to model MLC properties within the treatment planning system (TPS) for dynamic treatments. One of the key MLC properties in the Eclipse TPS is the dosimetric leaf gap (DLG) and precise determination of this parameter is paramount to ensuring accurate dose delivery. In this investigation, we report on how the spacing between control ...

متن کامل

On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans

Partial transmission through rounded leaf ends of Varian multileaf collimators (MLC) is accounted for with a parameter called the dosimetric leaf gap (DLG). Verification of the value of the DLG is needed when the dose delivery is accompanied by gantry rotation in VMAT plans. We compared the doses measured with GAFCHROMIC film and an ionization chamber to treatment planning system (TPS) calculat...

متن کامل

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

 ABSTRACTBackground: Although Electronic Portal imaging Devices (EPIDs) have originally developed for positioning verification, they can also be used for dosimetric purposes. In the current work, the dose response of minimum detectable thickness of a Scanning Liquid filled Ion Chamber EPID, SLIC-EPID, and the variation of transmitted dose with the shift of inhomogeneity inside of phantom was al...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2012