A novel phantom and procedure providing submillimeter accuracy in daily QA tests of accelerators used for stereotactic radiosurgery*
نویسندگان
چکیده
Stereotactic radiosurgery (SRS) places great demands on spatial accuracy. Steel BBs used as markers in quality assurance (QA) phantoms are clearly visible in MV and planar kV images, but artifacts compromise cone-beam CT (CBCT) isocenter localization. The purpose of this work was to develop a QA phantom for measuring with sub-mm accuracy isocenter congruence of planar kV, MV, and CBCT imaging systems and to design a practical QA procedure that includes daily Winston-Lutz (WL) tests and does not require computer aid. The salient feature of the phantom (Universal Alignment Ball (UAB)) is a novel marker for precisely localizing isocenters of CBCT, planar kV, and MV beams. It consists of a 25.4mm diameter sphere of polymethylmetacrylate (PMMA) containing a concentric 6.35mm diameter tungsten carbide ball. The large density difference between PMMA and the polystyrene foam in which the PMMA sphere is embedded yields a sharp image of the sphere for accurate CBCT registration. The tungsten carbide ball serves in finding isocenter in planar kV and MV images and in doing WL tests. With the aid of the UAB, CBCT isocenter was located within 0.10 ± 0.05 mm of its true positon, and MV isocenter was pinpointed in planar images to within 0.06 ± 0.04mm. In clinical morning QA tests extending over an 18 months period the UAB consistently yielded measurements with sub-mm accuracy. The average distance between isocenter defined by orthogonal kV images and CBCT measured 0.16 ± 0.12 mm. In WL tests the central ray of anterior beams defined by a 1.5 × 1.5 cm2 MLC field agreed with CBCT isocenter within 0.03 ± 0.14 mm in the lateral direction and within 0.10 ± 0.19 mm in the longitudinal direction. Lateral MV beams approached CBCT isocenter within 0.00 ± 0.11 mm in the vertical direction and within -0.14 ± 0.15 mm longitudinally. It took therapists about 10 min to do the tests. The novel QA phantom allows pinpointing CBCT and MV isocenter positions to better than 0.2 mm, using visual image registration. Under CBCT guidance, MLC-defined beams are deliverable with sub-mm spatial accuracy. The QA procedure is practical for daily tests by therapists.
منابع مشابه
A quality assurance method with submillimeter accuracy for stereotactic linear accelerators
The Stereotactic Alignment for Linear Accelerator (S. A. Linac) system is developed to conveniently improve the alignment accuracy of a conventional linac equipped with stereotactic cones. From the Winston-Lutz test, the SAlinac system performs three-dimensional (3D) reconstruction of the quality assurance (QA) ball coordinates with respect to the radiation isocenter, and combines this informat...
متن کاملDevelopment of a prototype stereotactic collimation assembly for Neptun 10 PC linac
Background: A set of hardware stereotactic radiosurgery is designed and constructed. The scope of this article is to describe how we have designed and developed stereotactic collimation assembly for a 9 MV Neptun 10 PC linac. Materials and methods: One collimator holder and six collimators were made, with their radiation field diameter between 12.5 mm to 25 mm at isocenter level. Beam data, suc...
متن کاملDesign, Construction and Evaluation of an Anthropomorphic Head Phantom for Assessment of Image Distortion in Stereotactic Radiosurgery Planning Systems
Introduction: In recent years, the use of magnetic resonance (MR) images in radiation treatment planning has drawn considerable attention. However, although the extent of a tumor can be determined in great detail on MR images, the geometric accuracy of these images is limited by distortions stemming from the inhomogeneity of the static background magnetic field, the nonlineari...
متن کاملAccurate Proton Beam Localization
The goal of this work was to further improve an experimental proton radiosurgery system at Loma Linda University Medical Center to reach sub-millimeter accuracy before proton radiosurgery with narrow beams can be used in a clinical trial. Radiosurgery precisely targets a specific anatomical region with high doses of radiation. We have developed a program that provides correcting translational o...
متن کاملDevelopment of A Head Docking Device for Linac-based Stereotactic Radiosurgery System to be Installed on Neptun 10 PC Linac
Introduction: Stereotactic radiosurgery is a technique for the treatment of intracranial lesions with highly collimated x-ray beam. Linac-based radiosurgery is currently performed by two techniques: couch or pedestal mounted. In the first technique, a device is required to affix patient's head to a couch and moreover to translate it accurately. Structure of suc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 17 شماره
صفحات -
تاریخ انتشار 2016