An implantable transducer for measuring tension in an anterior cruciate ligament graft.

نویسندگان

  • C P Ventura
  • J Wolchok
  • M L Hull
  • S M Howell
چکیده

The goal of this study was to develop a new implantable transducer for measuring anterior cruciate ligament (ACL) graft tension postoperatively in patients who have undergone ACL reconstructive surgery. A unique approach was taken of integrating the transducer into a femoral fixation device. To devise a practical in vivo calibration protocol for the fixation device transducer (FDT), several hypotheses were investigated: (1) The use of a cable versus the actual graft as the means for applying load to the FDT during calibration has no significant effect on the accuracy of the FDT tension measurements; (2) the number of flexion angles at which the device is calibrated has no significant effect on the accuracy of the FDT measurements; (3) the friction between the graft and femoral tunnel has no significant effect on measurement accuracy. To provide data for testing these hypotheses, the FDT was first calibrated with both a cable and a graft over the full range of flexion. Then graft tension was measured simultaneously with both the FDT on the femoral side and load cells, which were connected to the graft on the tibial side, as five cadaver knees were loaded externally. Measurements were made with both standard and overdrilled tunnels. The error in the FDT tension measurements was the difference between the graft tension measured by the FDT and the load cells. Results of the statistical analyses showed that neither the means of applying the calibration load, the number of flexion angles used for calibration, nor the tunnel size had a significant effect on the accuracy of the FDT. Thus a cable may be used instead of the graft to transmit loads to the FDT during calibration, thus simplifying the procedure. Accurate calibration requires data from just three flexion angles of 0, 45, and 90 deg and a curve fit to obtain a calibration curve over a continuous range of flexion within the limits of this angle group. Since friction did not adversely affect the measurement accuracy of the FDT, the femoral tunnel can be drilled to match the diameter of the graft and does not need to be overdrilled. Following these procedures, the error in measuring graft tension with the FDT averages less than 10 percent relative to a full-scale load of 257 N.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Static and fatigue strength of a fixation device transducer for measuring anterior cruciate ligament graft tension.

To determine which exercises do not overload the graft-fixation complex during intensive rehabilitation from reconstructive surgery of the anterior cruciate ligament (ACL), it would be useful to measure ACL graft loads during rehabilitative activities in vivo in humans. A previous paper by Ventura et al. (1998) reported on the design of an implantable transducer integrated into a femoral fixati...

متن کامل

In vivo calibration of a femoral fixation device transducer for measuring anterior cruciate ligament graft tension: a study in an ovine model.

Toward developing a transducer for measuring in vivo tension in anterior cruciate ligament grafts in humans, the objectives of this study were to determine the following: (1) whether the calibration of a previously reported femoral fixation device transducer (FDT) (Ventura et al., 1998) is affected by the presence of the graft when implanted in the tibial metaphysis of an ovine model, (2) wheth...

متن کامل

Anterior Cruciate Ligament Graft

Research Assistant, Biomedical Engineering Graduate Group. The goal of this study was to develop a new implantable transducer for measuring anterior cruciate ligament (ACL) graft tension postoperatively in patients who have undergone ACL reconstructive surgery. A unique approach was taken of integrating J. Wolcllok the transducer into a femoral fixation device. To devise a practical in vivo cal...

متن کامل

Femoral Condyle Fracture during Revision of Anterior Cruciate Ligament Reconstruction: Case Report and a Review of Literature

 Background: A rare and devastating complication following anterior cruciate ligament (ACL) revision reconstruction is femoral fracture.    Case presentation: A 35-year old male soccer player with a history of ACL tear from the previous year ago, who underwent arthroscopic ACL reconstruction and functioned well until another similar injury caused ACL re-rupture. Revision of ACL reconstruction w...

متن کامل

Contact pressure and tension in anterior cruciate ligament grafts subjected to roof impingement during passive extension.

Contact between an anterior cruciate ligament graft and the intercondylar roof has been termed roof impingement. Grafts with impingement sustain permanent damage, and if the injury is extensive enough, then the graft may fail, causing recurrent instability. This study evaluated two mechanical factors that could be responsible for the graft injury associated with roof impingement: an increase in...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of biomechanical engineering

دوره 120 3  شماره 

صفحات  -